Clinical Science (2008) 115, 233–244 (Printed in Great Britain) doi:10.1042/CS20070348

R

E

V

I

E

W

Diaphragm electromyography using an oesophageal catheter: current concepts Yuan Ming LUO∗ , John MOXHAM† and Michael I. POLKEY‡ *Guangzhou Medical College, National Key Laboratory of Respiratory Disease, Guangzhou 510210, People’s Republic of China, †Department of Respiratory Medicine, King’s College Hospital, London SE5 9PJ, U.K., and ‡Respiratory Muscle Laboratory, Royal Brompton Hospital, London SW3 6NP, U.K.

A

B

S

T

R

A

C

T

The usefulness of diaphragm electromyography recorded from an oesophageal electrode depends on a reliable signal which is free of artefact. The diaphragm EMG (electromyogram) recorded from chest wall surface electrodes may be unreliable because of signal contamination from muscle activity other than the diaphragm. Initially, the oesophageal electrode catheter for human studies had only one electrode pair, which could be difficult to position accurately and was influenced by a change in lung volume. Recently, a multipair oesophageal electrode has been developed which allows a high-quality EMG to be recorded. In the present review, the progress of oesophageal electrode design is outlined. The effects of signal contamination, electrode movement and particularly the effect of change in lung volume on the diaphragm EMG are discussed. The diaphragm EMG, recorded from a multipair oesophageal electrode, is useful to assess neural respiratory drive and diaphragm function in different groups of patients with respiratory disease, including patients with neuromuscular disease and sleep-disordered breathing, and those in the intensive care unit. When combined with cervical and cranial magnetic stimulation, an oesophageal electrode can be used to partition the central respiratory response time and phrenic nerve conduction time.

INTRODUCTION The diaphragm is the most important respiratory muscle, accounting for 70 % of respiration in normal humans [1]. Diaphragm weakness is associated with breathlessness and respiratory failure [2–6]. Electrical assessment of diaphragm and phrenic nerve function can provide important information on the diagnosis and management of patients with neuromuscular disease [7–13]. By recording the diaphragm CMAP (compound muscle action potential) and MEPs (motor-evoked potentials), respiratory neural pathways can also be assessed [14–19]. The diaphragm EMG (electromyogram) is also useful in detecting diaphragm fatigue [20–25] and assessing neural respiratory drive [26–34], as well as to trigger [33] and adjust

[34] mechanical ventilation. Measurement of the electrical activity offers different and complimentary information to that provided by the measurement of oesophageal pressure and Pdi (transdiaphragmatic pressure). However, the usefulness of the diaphragm EMG is dependent on the accurate recording of the signal without artefact. Three methods can be used in humans to record a diaphragm EMG: needles, surface electrodes and an oesophageal catheter. Although needle electrodes have been used in humans in small physiological studies [35,36], they are impractical for most clinical studies, particularly if there are factors which increase the risk of and from a pneumothorax. Chest wall surface electrodes provide a non-invasive technique and have been used to record the diaphragm

Key words: diaphragm, diaphragm function, electromyography, neural respiratory drive, oesophageal electrode, respiratory muscle. Abbreviations: CMAP, compound muscle action potential; COPD, chronic obstructive pulmonary disease; EMG, electromyogram; Fc, centroid frequencies; H/L ratio, ratio of high-frequency power to low-frequency power; ICU, intensive care unit; MEP, motorevoked potential; OSA, obstructive sleep apnoea; Pdi , transdiaphragmatic pressure; RMS, root mean square. Correspondence: Professor Michael I. Polkey (email [email protected]).  C

The Authors Journal compilation

 C

2008 Biochemical Society

233

234

Y. M. Luo, J. Moxham and M. I. Polkey

Figure 1 CMAP recorded simultaneously from lower chest wall and oesophageal electrodes during stimulation of the phrenic nerve only (upper traces) and both the brachial plexus and the phrenic nerve (lower traces)

In the lower chest wall recording, the latency of CMAP is shorter when stimulation co-activates the brachial plexus. The amplitude of CMAP is also reduced. In the oesophageal recordings, the waveforms of CMAP are similar. Four stimulations are superimposed; results are from one subject. This Figure was reproduced from Thorax (1999), volume 54, pp. 765–770, with permission from the BMJ Publishing Group. EMG, but this approach has disadvantages. First, the diaphragm EMG recorded from the chest wall surface electrodes can be unreliable because of signal contamination [37–40]. It has been reported that the diaphragm CMAP recorded from chest wall surface electrodes is likely to be contaminated by signals from adjacent muscles when the brachial plexus is co-activated, particularly likely with magnetic stimulation [37–39] (Figure 1). Similarly, the power spectrum and RMS (root mean square) activity derived from the spontaneous diaphragm EMG recorded from chest wall surface electrodes can be unreliable because of cross-talk signals from intercostal and abdominal muscles, which are also active during loaded breathing [40]. Secondly, the diaphragm EMG recorded from chest wall surface electrodes can be affected by subcutaneous fat, which significantly reduces signal strength due to muscle-to-electrode filtering effects [41]. Thirdly, it has been reported that the surface diaphragm CMAP is subject to power line artefacts [42]. Finally, although several possible sites have been reported [13,17,18,37–39,43,44], there is no standardized method for placing the chest wall electrodes, which makes data comparison between subjects and studies difficult. In contrast, the diaphragm EMG recorded from an oesophageal electrode is less affected by obesity, power line artefact and cross-talk signals [37–41]. Consequently, oesophageal diaphragm electromyography has become  C

The Authors Journal compilation

 C

2008 Biochemical Society

an increasingly important clinical and research technique [7,10,11,18,20–28,30–34,37–41,44]. Because assessment of diaphragm function, including electromechanical efficiency, is best done by recording both Pdi (gastric pressure−oesophageal pressure) and the diaphragm EMG, a single EMG electrode catheter mounted with two balloons for measurement of oesophageal and gastric pressure [45–47] has additional advantages over chest wall surface electrodes. Historically, the diaphragm was considered as one muscle [48]; however, studies have suggested that the diaphragm consists of two separate muscles, the costal and crural diaphragm [49]. This distinction is important because the oesophageal EMG measures signals from the crural component, whereas surface electrodes measure signals from the costal component. Some studies have found a small difference between the crural and costal diaphragm. An animal study [50] showed that there was more post-inspiratory activity in the costal than the crural muscle, but the distinction was most apparent during anaesthesia. It has also been reported, in cats, that the crural EMG is more sensitive to CO2 stimulation and to changes brought about by posture than the costal EMG [51]. However, the close correlation of the electrical activity in the crural and the costal parts has been well documented in humans during respiratory tasks [52,53]. The EMG recorded from the costal diaphragm was similar to that from the crus in terms of synchronous activation and time course in both quiet and loaded breathing [48,51,54]. Lastly, in a series of studies, Sharshar et al. [18] showed that the effect of various respiratory tasks on the diaphragm MEP was the same whether measured from oesophageal or surface electrodes. It should be noted, however, that other tasks relating to specific functions (e.g. postural or gastro-oesophageal sphincter control) may well be associated with differential activation of different portions of the diaphragm. The diaphragm EMG recorded by an oesophageal electrode is mainly from the crural part [55–58].

OESOPHAGEAL ELECTRODE DESIGN Oesophageal electrodes have been used for more than 40 years since they were introduced for recording human diaphragm EMGs in the 1960s [55,59]. An oesophageal electrode consists of an oesophageal catheter with metal coils at the distal end. Initially, oesophageal electrode catheters contained only one electrode pair [55,59]. Later, a modified oesophageal electrode with three pairs of electrodes was used [56]. Multipair electrodes with more than three pairs have been widely used [7,10,11,18,27,28,30–34,37–39,45,60,61]. Different metals have been used to construct oesophageal electrodes, including stainless steel [60,61], copper [23,25–28], silver [55,58,62] and platinum [46,63].

Oesophageal diaphragm electromyography

a good quality diaphragm EMG or CMAP can also be recorded using an interelectrode distance of 3 cm within a pair [7,10,11,27,28,38]. A power line artefact is one of the most common problems with recording the diaphragm EMG; it originates from the power line and mains power equipment and will typically have a 50 or 60 Hz signal depending on the power source. Proper earth connections and using a differential amplifier with a high common mode rejection ratio (e.g. > 100 dB) are useful to reduce the artefact [7,66]. An isolated pre-amplifier with ‘notch filters’ and shielded electrode cable can minimize the artefact further. Another source of artefact is electrode motion, which usually occurs during voluntary contractions, but it is relatively easy to deal with by using a high-pass-filter setting at 20 Hz [25,66]. The diaphragm CMAP produced by single twitch is usually free of motion artefact, as electrical activity occurs prior to any mechanical response [39]. Oesophageal peristalsis produces an additional lowfrequency large amplitude deflection of the baseline, but does not cause major problems if a high-pass-filter setting at 10 or 20 Hz is used [27,28,66]. Figure 2 Diaphragm CMAP recorded from an oesophageal electrode with different configurations

Left-hand panel, oesophageal electrode: pair A (electrodes 6 and 5), pair B (electrodes 6 and 4), pair C (electrodes 6 and 3), pair D (electrodes 6 and 2) and pair E (electrodes 6 and 1). Right-hand panel, diaphragm CMAP simultaneously recorded from the five pairs of electrodes during bilateral magnetic phrenic nerve stimulation. CMAP recorded from pair A has the smallest amplitude. The amplitude of CMAP increases with increasing interelectrode distance. The amplitude of CMAP recorded from pair C is 94 % of the amplitude of pair E. CMAP recorded from pair C has a more stable baseline than that recorded from pair D and E. Two signals are superimposed.

Interelectrode distance is important for the accuracy of recording the diaphragm EMG. Because the amplitude of the diaphragm CMAP elicited by phrenic nerve stimulation, and the magnitude of the diaphragm EMG during spontaneous breathing, relates to interelectrode distance [58] (Figure 2), to ensure data are comparable a constant interelectrode distance should be used. However, different interelectrode distances have been used since oesophageal recording was introduced. Initially, a 1 cm interelectrode distance was usually used [46,53,55,59,63], but other electrode configurations have also been employed; for example Gross et al. [64] used a 1.8 cm interelectrode distance. Some investigators have used a < 1 cm interelectrode distance [23,25,60,61]. McKenzie and Gandevia [58] found that the amplitude of CMAP increased with increasing interelectrode distance and they therefore used a 5 cm interelectrode distance [58,65]. There is no universal agreement about the optimal interelectrode distance, although an interelectrode distance of 1 cm was usually used to record interference pattern EMGs by Sinderby’s group [32,34]. We have found that

ECG ARTIFACT The diaphragm CMAP recorded from the oesophageal electrode can be contaminated by the superimposed ECG. The diaphragm EMG frequency is from 20– 250 Hz, whereas most of the ECG frequency is 0– 100 Hz [25]. It is therefore difficult to eliminate the ECG effectively using filters because of the overlap in the range of the two signals, although the P and T waves are relatively easy to delete by setting a high-pass filter at 20 Hz [20,25]. Wiener filtering could be useful to reduce the ECG artefact, but it is usually unable to completely separate the ECG from the EMG [67]. Many methods have been used to reduce or eliminate the ECG artefact. Gating techniques are often used online or offline, which delete both the EMG and QRS complex simultaneously for approx. 0.4 s around the QRS complex [25,68]. Some investigators have attempted to zero the EMG signal for the duration of QRS complex and to replace it with adjacent EMG activity [67] (Figure 3). A similar method, which we have favoured, is to analyse the EMG between QRS complexes [27,28,64], for example selected data between 50–75 % of the QRS complex interval [67], or to trigger sampling of the EMG with the ECG after a time delay [20]. Because the diaphragm EMG amplitude can be larger than the ECG during maximal breathing, it is not always reliable to trigger sampling with the ECG when using an oesophageal electrode, so surface signals are preferred for this purpose [20]. An additional channel for recording the ECG is sometimes necessary to act as a trigger source [20]. A more complex method is subtraction of the ECG template from the EMG signal in the  C

The Authors Journal compilation

 C

2008 Biochemical Society

235

236

Y. M. Luo, J. Moxham and M. I. Polkey

Figure 3 Diaphragm EMG contaminated by an ECG (lefthand panel) and the section with the ECG deleted using a gate method and replaced with a section from an adjacent EMG (right-hand panel)

section contaminated by the ECG waveform [68]. Several procedures are employed, including generating an ECG template by averaging the number of QRS complexes during expiration, and alignment of an ECG template to the ECG complexes overlapping the EMG and subtracting the ECG template from the EMG contaminated by ECG. These methods are considered to be more complicated than gating but also more accurate [68]. Fortunately, the effect of the ECG on the diaphragm CMAP elicited by phrenic nerve stimulation can be eliminated easily by simply discarding those CMAPs contaminated by the ECG, because the baseline and shape of the CMAP overlapped by the ECG will change [7,37–39,69].

EFFECT OF CHANGES IN LUNG VOLUME ON THE DIAPHRAGM EMG Concerns about changes in lung volume on the diaphragm EMG have hampered the application of the oesophageal diaphragm EMG in clinical practice. Many studies [53,60,62,65,70,71] have tried to address the effect of lung volume on the amplitude of the diaphragm EMG by using different electrode design. It was hypothesized that a change in the amplitude of the diaphragm CMAP was due to alternations in the distance between the electrode and the diaphragm secondary to the change in lung volume during respiration [65]. On this basis, a balloonanchored oesophageal electrode catheter was considered, at this time, to be useful to more reliably record the diaphragm CMAP. However, Smith and Bellemare [62] found that diaphragm CMAP changed with the movement of the electrode during respiration, despite using a balloon-anchored oesophageal catheter. To retain the anatomical relationship between the electrode and diaphragm during respiration, Gandevia and McKenzie  C

The Authors Journal compilation

 C

2008 Biochemical Society

[65] added an additional weight at the proximal part of the catheter, which itself had a ‘stabilized’ balloon at the distal end, to let the catheter be slightly pulled out by the weight during expiration and pulled in by the pushing of the balloon by the diaphragm during inspiration. However, they found the diaphragm CMAP amplitude (peak to peak) at TLC (total lung capacity) was 3.4 times larger than at RV (residual volume), indicating either that their technique could not effectively maintain electrode position or that the distance between the electrode and the diaphragm was not the cause of the variance. They also observed that during deep breathing the electrode catheter could move up to 8 cm, which is beyond the maximal movement of the crus [61,70]. The large movement of the catheter during respiration suggested that the catheter with both a balloon and the weight was actually unable to reliably record the diaphragm EMG because the electrodes failed to track the movement of the crus of the diaphragm [70,71]. Previously, we have used a multipair oesophageal electrode with an interelectrode distance of 3 cm within a pair and found that changes in lung volume of up to 2 litres had little effect on the amplitude of the diaphragm CMAP, providing the oesophageal electrode catheter was optimally positioned and then fixed on the nose [71]. However, Beck et al. [70] reported that the amplitude of the diaphragm CMAP changes with lung volume when using an electrode with an interelectrode distance of 1.0 cm within a pair. Nevertheless, it now seems clear that the oesophageal electrode with a stabilized balloon does not record the diaphragm EMG reliably [27,28,56,61,70]. To overcome the influence of lung volume on the diaphragm EMG, multipair oesophageal electrodes have been used to record the diaphragm EMG. Daubenspeck et al. [60] found that the diaphragm EMG could be reliably recorded by summation of the EMG recorded from seven consecutive pairs of electrode in spite of a change in lung volume. It has been suggested that the orientation of the crural diaphragm fibres with respect to the axis of the oesophagus is perpendicular [41]. Therefore the electrode pairs caudal and cephalad to the central area of the crus would have a similar amplitude and opposite polarity if both electrode pairs were the same distance from the centre [7,38,39,71]. Sinderby and co-workers [67] have developed a cross-correlation method to determine the electrode pair closest to the diaphragm. A catheter with seven [72] or eight [67] consecutive pairs of electrodes with a 1 cm interelectrode distance within a pair was used to track the position of the diaphragm’s electrically active centre. A correlation coefficient was calculated for every other pair and the difference between the signals that had the best correlation was obtained (the doublesubtracted signal) [67,72] to enhance the signal-tonoise ratio and to reduce the artefact originating from diaphragm movement. With this technique, they also found that the diaphragm EMG was independent of changes in lung volume; however, it should be pointed out

Oesophageal diaphragm electromyography

that the double-subtracted signal was from the electrode pairs slightly away from the diaphragm, rather than the pair at the electrically active centre, although the signal obtained from the pair exactly at the electrical activity centre was sometimes included for analysis [45].

POSITIONING OF ELECTRODES To accurately record the diaphragm EMG, the oesophageal electrode should be positioned as close to the diaphragm as possible. Unlike chest wall surface electrodes, oesophageal electrodes cannot be positioned simply using fixed anatomical landmarks. Traditionally, investigators [23] positioned the oesophageal electrode by observing the magnitude of the diaphragm EMG during sniffs or during maximal or quiet breathing efforts on the assumption that the maximal EMG signal would be obtained when the catheter was optimally positioned [23,55,73]. Luo et al. [7,38,71] developed a technique to accurately position an electrode based on the amplitude and polarity of the diaphragm CMAP elicited by bilateral magnetic phrenic nerve stimulation [7,71]. We used a catheter with seven metal coils (number 1 being proximal and 7 being distal) created with four electrode pairs (Figure 4) , the interelectrode distance within a pair being 3 cm. Electrode 3 was positioned close to the diaphragm based on the amplitude and opposite polarity of the diaphragm CMAP recorded from pairs 1 and 3, since the CMAP shape is sensitive to electrode position [61,74]. The position was adjusted further by observing the amplitude and polarity of the CMAP recorded from pair 2, which was characterized by a small or absent CMAP, since there is a cancellation of the potential when both recording electrodes are equally close to the source of the diaphragm potential [65,71,72]. The ideal position of the electrode was characterized by a large CMAP amplitude from pairs 1 and 3, with opposite polarity, and a small CMAP was recorded from electrode pair 2 (Figure 4). This method can position the electrode accurately, but it requires the use of phrenic nerve stimulation which is a complex technique. Luo et al. [27,28] have demonstrated the positioning of a multipair oesophageal electrode based on a spontaneous EMG recorded simultaneously from four or five pairs of electrodes. The optimal position was characterized by a large EMG from two electrode pairs which shared the middle electrode and a small EMG from the pair straddling the middle electrode (Figure 5).

Figure 4 Diaphragm CMAP recorded from four pairs of electrodes

Left-hand panel, an oesophageal electrode catheter consisting of seven coils which form four recording electrode pairs. Each electrode is 1 cm in length and there is a 1 cm gap between electrodes. The interelectrode distance within an electrode pair is 3 cm. Electrode 3 is placed at the centre of the electrically active region of the diaphragm (EARdi). Right-hand panel, CMAP recorded from four electrode pairs at functional residual capacity. The signal from pairs one and three are opposite in polarity. Pair 2 only records a small CMAP since both electrodes are close to the EARdi centre. Pair 4 records a small CMAP, as the electrode is 3 cm away from the diaphragm. Reproduced from [71], with permission. Copyright (2000) European Respiratory Society Journals Ltd. by the stimulator output. A supramaximal stimulation of the phrenic nerve is usually desirable, although since submaximal stimuli recruit large (and therefore faster) bundles of axons preferentially the measured nerve conduction time does not greatly depend on the stimulus intensity whereas the amplitude does (Figure 6). For the amplitude of the diaphragm CMAP to be useful, only the signals elicited by demonstrably supramaximal (conventionally 20 % higher than maximal) stimulation are usually analysed. To eliminate the influence of the ECG, CMAPs are usually selected from those with a constant shape and stable baseline before and after stimulation [7,38,39,69]. Similar principles are used in the analysis of the MEP [14–19] following stimulation of the diaphragm motor area of the cortex, but it should be noted that, because the MEP reflects a volley of impulses, deciding where the signal begins (and hence the latency) may be more difficult (Figure 7).

TIME AND FREQUENCY DOMAIN ANALYSIS OF THE DIAPHRAGM [75]

ANALYSIS OF EMG SIGNALS

Time domain EMG analysis

For the diaphragm CMAP, latency and the amplitude are measured [7,14,37–39,44,58]. The latency is synonymous with the conduction time, which is measured from stimulation artefact to the onset of CMAP [7,14,37–39,44,58]. The amplitude of the diaphragm CMAP is greatly affected

Time domain EMG analysis is used to describe the relationship between EMG strength and time so that, broadly speaking, activity is the product of the amplitude of the signal and its duration. To facilitate computerized processing, the signal is usually rectified and integrated [46,63,76,77]; however, some groups prefer to use a  C

The Authors Journal compilation

 C

2008 Biochemical Society

237

238

Y. M. Luo, J. Moxham and M. I. Polkey

Figure 5 Diaphragm EMG recorded from a multipair oesophageal electrode during CO2 rebreathing

The oesophageal electrode configuration is shown in the left-hand panel. The oesophageal electrode catheter consists of ten coils that form five recording electrode pairs. Each electrode is 1 cm in length, and there is a 0.5 mm gap between electrodes. The interelectrode distance within an electrode pair is 3.2 cm. Electrode 5 is accurately positioned at the centre of the diaphragm electrically active region on the basis of the amplitude of the diaphragm EMG recorded from the five pairs of electrodes during resting spontaneous breathing. This is characterized by a large EMG from pairs 1 and 5 and the smallest EMG from pair 3. Pair 2 and pair 4 recorded a smaller EMG than pairs 1 and 5 because they were slightly further away from the diaphragm electrically active centre. Pair 3 only recorded the smallest EMG because of signal cancellation. The diaphragm EMG increases during CO2 rebreathing. In spite of the dynamic hyperventilation that occurs in patients with COPD, the electrode pairs which recorded the maximal EMG at rest (pairs 1 or 5) were the same as that at the end of CO2 rebreathing, indicating that lung volume change did not significantly affect electrode positions. This Figure was reprinted from Respiratory Physiology & Neurobiology , volume 146, Y.M. Luo and J. Moxham, Measurement of neural respiratory drive in patients with COPD, pp. 165–174, Copyright (2005), with permission from Elsevier (http://www.sciencedirect.com/science/journal/15699048). moving average, peak activity or rate of rise of activity [26]. The RMS of the diaphragm EMG has been used (Figure 8) to quantify the activity of the diaphragm [27,28,31,34,78] because it reflects the number and firing rate of the motor units recruited. It has been demonstrated that the RMS of the diaphragm EMG is independent of a change in lung volume and is a reliable index of global diaphragm activation [45,67,70].

of high-frequency power to low-frequency power) are usually measured [25]. Fc reflect the relative contribution of low- and high-frequency components in a given bandwidth, and its number represents that frequency at which the power above and below it is the same [23,25]. The H/L ratio is the ratio of power at high frequencies (129–246 Hz) to power at low frequencies (31–51 Hz) [25]. Both Fc and the H/L ratio have been used to identify diaphragm fatigue [20,21,23,25].

Frequency domain analysis Power spectral analysis of the EMG has been used to assess diaphragm fatigue [20,21,24,25,73,79], although enthusiasm for its use has diminished following the demonstration that EMG indices of fatigue correlate poorly with mechanical indices of fatigue [80] (Figure 9). In brief, this approach seeks to describe the relationship between the frequency and power of the diaphragm EMG. Fc (centroid frequencies) and the H/L ratio (ratio  C

The Authors Journal compilation

 C

2008 Biochemical Society

APPLICATIONS Testing the integrity of the phrenic nerve–diaphragm unit Accurately recording the diaphragm CMAP elicited by electrical or magnetic stimulation of the phrenic nerve is very useful for assessing diaphragm function and for

Oesophageal diaphragm electromyography

Figure 6 CMAP recorded from the oesophageal electrode elicited by electrical stimulation and by unilateral magnetic stimulation at different stimulator output

The amplitude of CMAP increases with increasing stimulator output, and becomes the same as with electrical stimulation. Reproduced from [38], with permission. Copyright (1999) European Respiratory Society Journals Ltd.

Figure 7 Examples of results produced by transcranial magnetic stimulation during a facilitated inspiration

Left-hand panel, MEP; right-hand panel, P di , oesophageal pressure (P oes) and gastric pressure (P gas). This Figure was published in Magnetic Stimulation in Clinical Neurophysiology, 2nd edn (Hallett, M. and Chokroverty, S., eds), Polkey, M.I. and Moxham, J., Magnetic stimulation in the assessment of the respiratory muscle pump, pp. 381–392, Copyright (2005), with permission from Elsevier.

selecting patients for diaphragm pacing. Because surface electrodes may be associated with varying conduction times [38,39,81] and are not practical in the obese or those with a chest wall deformity, the oesophageal electrode often offers a better chance of achieving accurate data. Measurement of phrenic nerve conduction time with an oesophageal electrode is useful in differentiating axonal

neuropathies (e.g. motor neuron disease) from neuropathies where damage primarily occurs to the axonal myelin [7–9,12,13,69]. Patients with Guillain-Barr´e syndrome and hereditary motor sensory type 1 usually have a prolongation of phrenic nerve conduction time [8,13]. In contrast, conduction times in patients with motor neuron disease or hereditary motor sensory type 2 are usually normal [13]. The normal range for phrenic nerve conduction time measured by an oesophageal electrode is 7.6 + − 0.7 ms on the left and 6.9 + − 0.9 ms on the right [7]. This reflects the length of the nerve, with the phrenic nerve conduction time on the right side being slightly, but significantly, shorter than that on the left [58]. Oesophageal electrodes can also be used to assess patients who have undergone open chest surgery, including cardiac surgery, to determine whether the phrenic nerve has been injured during the operation [9]. Because the multipair oesophageal electrode can reliably record the diaphragm CMAP elicited by magnetic stimulation, patients with suspected hemi-diaphragm or bilateral diaphragm paralysis can be accurately diagnosed [7,10], although the assessment of the magnitude weakness will usually require the measurement of twitch Pdi [7,82].

Assessment of cortical diaphragmatic pathways Oesophageal electrodes can be used to record the diaphragm EMG response to transcranial electrical [83] or magnetic [14–19] stimulation to investigate neural pathways of the diaphragm. Sharshar et al. [18] found that the diaphragm MEP could be reliably recorded with an oesophageal electrode (Figure 7). By combining transcranial and cervical magnetic stimulation, central conduction time can be estimated by subtraction of the latency of the CMAP elicited by neck stimulation from transcranial stimulation [14–19,84]. On the basis of the MEP, it has been suggested that patients with stroke have impaired central motor conduction to the diaphragm [17]. A possible supplementary diaphragm motor area (3 cm anterior to the vertex) has been identified during transcranial magnetic stimulation mapping with the help of an oesophageal electrode [85]. Motor control of the costal and crural diaphragm can also been assessed with the help of the oesophageal EMG technique [18].

Efficiency of diaphragm contraction and diaphragm fatigue The EMG recorded from an oesophageal electrode can be used to establish the relationship between EMG amplitude and Pdi . Similar to the relationship between quadriceps force and EMG amplitude [86], a linear relationship exists between Pdi and the diaphragm EMG recorded from needle [87] or oesophageal [88] electrodes.  C

The Authors Journal compilation

 C

2008 Biochemical Society

239

240

Y. M. Luo, J. Moxham and M. I. Polkey

Figure 8 Recording and analysis of the diaphragm EMG

The diaphragm EMG is recorded with an oesophageal catheter with ten coils. EMG signals are amplified with a gain of 1000 and filtered. RMS of the diaphragm EMG is generated after eliminating the ECG.

Figure 9 Force response to electrical stimulation at 20 Hz/force at 100 Hz following repeated submaximal voluntary contractions in the quadriceps

The Figure illustrates substantial low-frequency fatigue, with the EMG H/L ratio remaining normal. This Figure was reproduced from Moxham, J., Edwards, R.H.T., Aubier, M. et al. (1982) Changes in EMG power spectrum (high/low ratio) with force fatigue in man, Journal of Applied Physiology , 53, 1094–1099, and is used with permission. Copyright American Physiological Society.

It has also been shown that there is a linear relationship between twitch Pdi and the amplitude of the diaphragm CMAP (r = 0.8) [7], which could be clinically important because the diaphragm EMG could be used to predict the strength of the diaphragm, assuming neuromuscular transmission is normal. When Pdi and the diaphragm EMG are measured simultaneously, contraction effi C

The Authors Journal compilation

 C

2008 Biochemical Society

ciency can be assessed and peripheral low-frequency diaphragm fatigue can be detected based on a decrease in the Pdi /integrated diaphragm EMG ratio [89]. The power spectrum of the diaphragm EMG can be recorded from an oesophageal electrode and a shift to lower frequency has been considered to be an indicator that fatigue is occurring [20–25,64,73,79,80]. Fc and the H/L ratio have been used to diagnose diaphragm fatigue in patients, including those with OSA (obstructive sleep apnoea) [24,79], patients undergoing weaning from mechanical ventilation [21] and patients with COPD (chronic obstructive pulmonary disease) during exhaustive exercise [23,32]. From power spectral data, it has been considered that diaphragm fatigue contributes to weaning failure and respiratory symptoms [3,6,90], but findings obtained using magnetic nerve stimulation have established that true low-frequency fatigue is hard to elicit in clinical practice [91–94]. Because the relationship between the power spectral shift and fatigue remains controversial and the underlying cellular mechanisms remain unknown, in particular because power spectral analysis can be affected by the change in lung volume and the distance between the electrode and the diaphragm [70,72], the shift of the power spectrum to lower frequency is of uncertain clinical significance.

Assessment of neural respiratory drive It is increasingly useful to assess neural respiratory drive, as the product of EMG amplitude against time, in normal subjects and patients with respiratory disease. The diaphragm EMG has long been considered to be a powerful physiological tool to assess neural respiratory drive [26–34,52,56,71]. The diaphragm EMG increases

Oesophageal diaphragm electromyography

Figure 10 Peak of RMS increases during CO2 rebreathing

There is a good relationship between end-tidal CO2 and the amplitude of the RMS. No plateau of EMG is observed. Results are from a patient with COPD. This Figure was reprinted from Respiratory Physiology & Neurobiology , volume 146, Y.M. Luo and J. Moxham, Measurement of neural respiratory drive in patients with COPD, pp. 165–174, Copyright (2005), with permission from Elsevier (http://www.sciencedirect.com/science/journal/15699048). gradually during CO2 rebreathing [26–28,46,76], and there is a linear relationship between the RMS of the diaphragm EMG and ventilation in normal subjects [28] and patients with COPD [27] (Figure 10). The observation that pressure support reduces both Pdi and diaphragm EMG equally in normal subjects [34] suggests that two measures reflect neural respiratory drive. However, inspiratory pressure underestimates neural respiratory drive in patients with COPD because of hyperinflation. For example, Sinderby et al. [32] have shown that diaphragm electrical activity increased progressively during incremental exercise in COPD, whereas Pdi reached a plateau, indicating that the diaphragm EMG is a better index of neural respiratory drive. Luo and Moxham [27] found neural drive, as assessed by RMS of the diaphragm EMG, increased linearly with an increase in end-tidal CO2 in patients with COPD. Furthermore, they found that the diaphragm EMG in patients with COPD increased initially but eventually reached a plateau during constant rate treadmill exercise, indicating the importance of the exercise protocol.

Application in the ICU (intensive care unit) An important application of the oesophageal diaphragm EMG is in the assessment of diaphragm function in patients in the ICU [11]. Because patients in the ICU frequently develop a neuromuscular abnormality [12,95,96], including dysfunction of the diaphragm [97], which contribute to weaning failure, measurement of the diaphragm CMAP with an oesophageal electrode can provide useful information about phrenic nerve and dia-

Figure 11 Diaphragm EMG recorded from five pairs of oesophageal electrodes during polysomnography

Signals from top to bottom are two-channel EEGs, two-channel EOGs (electrooculograms), airflow recorded from pneumotachometer, EMGs (1–5) recorded from a multipair oesophageal electrode, chest and abdominal movement and S aO2 % (percentage arterial O2 saturation). During cessation of airflow, there is barely any chest and abdominal movement, therefore, the event appears to be central sleep apnoea. However, there is obvious EMG during cessation of the airflow, and the event should therefore be defined as OSA. The inconsistent changes between chest/abdominal movement and the diaphragm EMG suggests that the diaphragm EMG recorded from oesophageal electrode can more accurately identify the type of respiratory events. Results are from a patient with OSA. phragm function [7,8]. Luo et al. [11] measured CMAP with a multipair oesophageal electrode in the ICU and found that the amplitude of the diaphragm CMAP in patients was half that of normal subjects. An attractive potential application of the oesophageal electrode is triggering and adjusting ventilation from the diaphragm EMG [30–34]. Conventionally, ventilators are triggered by flow or pressure, but these methods could fail to detect the onset of breathing causing asynchrony between the patient and the ventilator [33]. Because the diaphragm EMG appears ahead of airflow or a change of pressure, particularly so when there is intrinsic or applied positive end-expiratory pressure, the diaphragm EMG recorded from a multipair oesophageal electrode could be used to trigger the ventilator and improve patient–ventilator synchrony [33]. The diaphragm EMG can also be used to automatically adjust a ventilator in response to varying patient effort [30,33,34]. With this model, ventilatory assistance is given in proportion to the neural drive. Spahija et al. [30] reported that, during loaded breathing in healthy subjects, it was possible to automatically adjust the level of ventilatory assistance from one breath  C

The Authors Journal compilation

 C

2008 Biochemical Society

241

242

Y. M. Luo, J. Moxham and M. I. Polkey

to the next and to maintain diaphragm activation within a predetermined target range. Ventilator adjustment using the diaphragm EMG can efficiently unload the respiratory muscles during maximal inspiration [34].

Applications in sleep medicine Vincken et al. [98] used an oesophageal electrode to investigate the mechanism of airway opening in OSA and concluded that reopening of the airway is not related to respiratory muscle fatigue. Some studies have investigated diaphragm activity during obstructive events in an attempt to determine whether OSA is associated with diaphragm derecruitment [24,79]. Conventionally, distinguishing obstructive from central apnoea/hypopnoea depends on detection of respiratory effort measured from chest and abdominal movement or oesophageal pressure. The diaphragm EMG can accurately assess neural respiratory drive [27,28,31–33,99], therefore, it could be a useful technique to differentiate types of respiratory events [100,101] (Figure 11); this distinction may be of importance in conditions such as heart failure, where nocturnal ventilatory support is indicated for patients with OSA but not central sleep apnoea.

CONCLUSIONS In conclusion, measuring the diaphragm EMG may be useful for the diagnosis and treatment of respiratory disease. Recent technical advances mean that an oesophageal electrode is simpler to use and, therefore, should now, in most cases, be the technique of choice in preference to surface electrodes.

ACKNOWLEDGMENTS Y. M. L. is supported by a grant from the Natural Science Foundation of Guangdong, People’s Republic of China (No. 6113858) and holds British Council researcher exchange programmes awards.

REFERENCES 1 Mead, J. and Loring, S. H. (1982) Analysis of volume displacement and length changes of the diaphragm during breathing. J. Appl. Physiol. 53, 750–755 2 Macklem, P. T. and Roussos, C. S. (1977) Respiratory muscle fatigue: a cause of respiratory failure? Clin. Sci. Mol. Med. 53, 419–422 3 Roussos, C. (1985) Function and fatigue of respiratory muscles. Chest 88, 124S–132S 4 Aubier, M.(1989) Respiratory muscle fatigue. Intensive Care Med. 15, S17–S20 5 Cohen, C. A., Zagelbaum, G., Gross, D., Roussos, C. and Macklem, P. T.(1982) Clinical manifestations of inspiratory muscle fatigue. Am. J. Med. 73, 308–316 6 Roussos, C., Fixley, M., Gross, D. and Macklem, P. T. (1979) Fatigue of the inspiratory muscles and their synergic behaviour. J. Appl. Physiol. 46, 897–904  C

The Authors Journal compilation

 C

2008 Biochemical Society

7 Luo, Y. M., Lyall, R. A., Harris, M. L., Rafferty, G. F., Polkey, M. I. and Moxham, J. (1999) Quantification of the esophageal diaphragm EMG with magnetic phrenic nerve stimulation. Am. J. Respir. Crit. Care Med. 160, 1629–1634 8 Gourie-Devi, M. and Ganapathy, G. R. (1985) Phrenic nerve conduction time in Guillain-Barr´e syndrome. J. Neurol. Neurosurg Psychiatry 48, 245–249 9 Mok, Q., Ross-Russell, R., Mulvey, D., Green, M. and Shinebourne, E. A. (1991) Phrenic nerve injury in infants and children undergoing cardiac surgery. Br. Heart J. 65, 287–292 10 Luo, Y. M., Harris, M. L., Lyall, R. A., Watson, A., Polkey, M. I. and Moxham, J. (2000) Assessment of diaphragm paralysis with oesophageal electromyography and unilateral magnetic phrenic nerve stimulation. Eur. Respir. J. 15, 596–599 11 Luo, Y. M., Chen, R. C. and Zhong, N. S. (2005) Measurement of diaphragm compound muscle action potential with magnetic stimulation of the phrenic nerve and multipair esophageal electrode in intensive care unit. Chin. J. Tuberc. Respir. Dis. 28, 505–508 12 Deem, S., Lee, C. M. and Curtis, J. R. (2003) Acquired neuromuscular disorders in the intensive care unit. Am. J. Respir. Crit. Care Med. 168, 735–739 13 Markand, O. N., Kincaid, J. C., Pourmand, R. A. et al. (1984) Electrophysiologic evaluation of diaphragm by transcutaneous phrenic nerve stimulation. Neurology 34, 604–614 14 Zifko, U., Remtulla, H., Power, K., Harker, L. and Bolton, C. F. (1996) Transcortical and cervical magnetic stimulation with recording of the diaphragm. Muscle Nerve 19, 614–620 15 Similowski, T., Straus, C., Attali, V., Duguet, A., Jourdain, B. and Derenne, J. P. (1996) Assessment of the motor pathway to the diaphragm using cortical and cervicalmagnetic stimulation in the decision-making process of phrenic pacing. Chest 110, 1551–1557 16 Similowski, T., Straus, C., Coic, L. and Derenne, J. P. (1996) Facilitation-independent response of the diaphragm to cortical magnetic stimulation. Am. J. Respir. Crit. Care Med. 154, 1771–1777 17 Similowski, T., Catala, M., Rancurel, G. and Derenne, J. P. (1996) Impairment of central motor conduction to the diaphragm in stroke. Am. J. Respir. Crit. Care Med. 154, 436–441 18 Sharshar, T., Hopkinson, N. S., Ross, E. T. et al. (2005) Motor control of the costal and crural diaphragm–insights from transcranial magnetic stimulation in man. Respir. Physiol. Neurobiol. 146, 5–19 19 Similowski, T., Duguet, A., Straus, C., Attali, V., Boisteanu, D. and Derenne, J. P. (1996) Assessment of the voluntary activation of the diaphragm using cervical and cortical magnetic stimulation. Eur. Respir. J. 9, 1224–1231 20 Bower, J. S., Sandercock, T. G., Rothman, E., Abbrecht, P. H. and Dantzker, D. R. (1984) Time domain analysis of diaphragmatic electromyogram during fatigue in men. J. Appl. Physiol. 57, 913–916 21 Brochard, L., Harf, A., Lorino, H. and Lemaire, F. (1989) Inspiratory pressure support prevents diaphragmatic fatigue during weaning from mechanical ventilation. Am. Rev. Respir. Dis. 139, 513–521 22 Grassino, A., Bellemare, F. and Laporta, D. (1984) Diaphragm fatigue and the strategy of breathing in COPD. Chest 85, 51S–54S 23 Aldrich, T. K., Adams, J. M., Arora, N. S. and Rochester, D. F. (1983) Power spectral analysis of the diaphragm electromyogram. J. Appl. Physiol. 54, 1579–1584 24 Montserrat, J. M., Kosmas, E. N., Cosio, M. G. and Kimoff, R. J. (1997) Lack of evidence for diaphragmatic fatigue over the course of the night in obstructive sleep apnoea. Eur. Respir. J. 10, 133–138 25 Schweitzer, T. W., Fitzgerald, J. W., Bowden, J. A. and Lynne-Davies, P. (1979) Spectral analysis of human inspiratory diaphragmatic electromyograms. J. Appl. Physiol. 46, 152–165 26 Lopata, M., Evanich, M. J. and Lourenco, R. V. (1977) Quantification of diaphragmatic EMG response to CO2 rebreathing in humans. J. Appl. Physiol. 43, 262–270

Oesophageal diaphragm electromyography

27 Luo, Y. M. and Moxham, J. (2005) Measurement of neural respiratory drive in patients with COPD. Respir. Physiol. Neurobiol. 146, 165–174 28 Luo, Y. M., Hart, N., Mustfa, N. et al. (2001) Effect of diaphragm fatigue on neural respiratory drive. J. Appl. Physiol. 90, 1691–1699 29 Lahrmann, H., Wild, M., Wanke, T. et al. (1999) Drive to the diaphragm after lung volume reduction surgery. Chest 116, 1593–1600 30 Spahija, J., Beck, J., De Marchie, M., Comtois, A. and Sinderby, C. (2005) Closed-loop control of respiratory drive using pressure-support ventilation: target drive ventilation. Am. J. Respir. Crit. Care Med. 171, 1009–1014 31 Beck, J., Gottfried, S.B., Navalesi, P. et al.(2001) Electrical activity of the diaphragm during pressure support ventilation in acute respiratory failure. Am. J. Respir. Crit. Care Med. 164, 419–424 32 Sinderby, C., Spahija, J., Beck, J. et al. (2001) Diaphragm activation during exercise in chronic obstructive pulmonary disease. Am. J. Respir. Crit. Care Med. 163, 1637–1641 33 Sinderby, C., Navalesi, P., Beck, J. et al. (1999) Neural control of mechanical ventilation in respiratory failure. Nat. Med. 5, 1433–1436 34 Sinderby, C., Beck, J., Spahija, J. et al. (2007) Inspiratory muscle unloading by neurally adjusted ventilatory assist during maximal inspiratory efforts in healthy subjects. Chest 131, 711–717 35 Demoule, A., Verin, E., Locher, C., Derenne, J. P. and Similowski, T. (2003) Validation of surface recordings of the diaphragm response to transcranial magnetic stimulation in humans. J. Appl. Physiol. 94, 453–461 36 Gorman, R. B., McKenzie, D. K., Butler, J. E., Tolman, J. F. and Gandevia, S. C. (2005) Diaphragm length and neural drive after lung volume reduction surgery. Am. J. Respir. Crit. Care Med. 172, 1259–1266 37 Luo, Y. M., Polkey, M. I., Lyall, R. A. and Moxham, J. (1999) Effect of brachial plexus coactivation on phrenic nerve conduction time. Thorax 54, 765–770 38 Luo, Y. M., Johnson, L. C., Polkey, M. I. et al. (1999) Diaphragm electromyogram measured with unilateral magnetic stimulation. Eur. Respir. J. 13, 385–390 39 Luo, Y. M., Polkey, M. I., Johnson, L. C. et al. (1998) Diaphragm EMG measured by cervical magnetic and electrical phrenic nerve stimulation. J. Appl. Physiol. 85, 2089–2099 40 Sinderby, C., Friberg, S., Comtois, N. and Grassino, A. (1996) Chest wall muscle cross talk in canine costal diaphragm electromyogram. J. Appl. Physiol. 81, 2312–2327 41 Beck, J., Sinderby, C., Weinberg, J. and Grassino, A. (1995) Effects of muscle-to-electrode distance on the human diaphragm electromyogram. J. Appl. Physiol. 79, 975–985 42 Armaganidis, A. and Roussos, C. (1995) Measurement of the work of breathing in the critically ill patient. The Thorax 2nd edn (Roussos, C., ed.), pp. 1231–1274, Marcel Dekker, New York 43 Glerant, J. C., Mustfa, N., Man, W. D. et al. (2006) Diaphragm electromyograms recorded from multiple surface electrodes following magnetic stimulation. Eur. Respir. J. 27, 334–342 44 Similowski, T., Mehiri, S., Duguet, A., Attali, V., Straus, C. and Derenne, J. P. (1997) Comparison of magnetic and electrical phrenic nerve stimulation in assessment of phrenic nerve conduction time. J. Appl. Physiol. 82, 1190–1199 45 Sinderby, C., Spahija, J., Beck, J. et al. (2001) Diaphragm activation during exercise in chronic obstructive pulmonary disease. Am. J. Respir. Crit. Care Med. 163, 1637–1641 46 Onal, E., Lopata, M., Ginzburg, A. S. and O’Connor, T. D. (1981) Diaphragmatic EMG and transdiaphragmatic pressure measurements with a single catheter. Am. Rev. Respir. Dis. 124, 563–565 47 Luo, Y. M. and Zhong, N. S. (1994) Diaphragm fatigue in patients with obstructive pulmonary disease. Chin. J. Tuberc. Respir. Dis. 17, 93–96

48 Sant’Ambrogio, G., Frazier, D. T., Wilson, M. F. and Agostoni, E. (1963) Motor innervation and pattern of activity of cat diaphragm. J. Appl. Physiol. 18, 43–46 49 De Troyer, A., Sampson, M., Sigrist, S. and Macklem, P. T. (1981) The diaphragm: two muscles. Science 213, 237–238 50 Henderson-Smart, D. J., Johnson, P. and McClelland, M. E. (1982) Asynchronous respiratory activity of the diaphragm during spontaneous breathing in the lamb. J. Physiol. 327, 377–391 51 Van Lunteren, E., Haxhiu, M. A., Cherniack, N. S. and Goldman, M. D. (1985) Differential costal and crural diaphragm compensation for posture changes. J. Appl. Physiol. 58, 1895–1900 52 Lourenco, R. V., Cherniack, N. S., Malm, J. R. and Fishman, A. P. (1966) Nervous output from the respiratory center during obstructed breathing. J. Appl. Physiol. 21, 527–533 53 Kim, M. J., Druz, W. S., Danon, J., Machnach, W. and Sharp, J. T. (1978) Effects of lung volume and electrode position on the esophageal diaphragmatic EMG. J. Appl. Physiol. 45, 392–398 54 Pollard, M. J., Megirian, D. and Sherrey, J. H. (1985) Unity of costal and crural diaphragmatic activity in respiration. Exp. Neurol. 90, 187–193 55 Agostoni, E., Sant’ambrogio, G. and Del Portillo Carrasco, H. (1960) Electromyography of the diaphragm in man and transdiaphragmatic pressure. J. Appl. Physiol. 15, 1093–1097 56 Onal, E., Lopata, M. and Evanich, M. J. (1979) Effects of electrode position on esophageal diaphragmatic EMG in humans. J. Appl. Physiol. 47, 1234–1238 57 Lourenco, R. V. and Mueller, E. P. (1967) Quantification of electrical activity in human diaphragm. J. Appl. Physiol. 22, 598–600 58 McKenzie, D. K. and Gandevia, S. C. (1985) Phrenic nerve conduction times and twitch pressures of the human diaphragm. J. Appl. Physiol. 58, 1496–1504 59 Petit, J. M., Milic-Emili, G. and Delhez, L. (1960) Role of the diaphragm in breathing in conscious normal man: an electromyographic study. J. Appl. Physiol. 15, 1101–1106 60 Daubenspeck, J. A., Leiter, J. C., McGovern, J. F., Knuth, S. L. and Kobylarz, E. J. (1989) Diaphragmatic electromyography using a multiple electrode array. J. Appl. Physiol. 67, 1525–1534 61 Beck, J., Sinderby, C., Lindstrom, L. and Grassino, A. (1996) Influence of bipolar esophageal electrode positioning on measurements of human crural diaphragm electromyogram. J. Appl. Physiol. 81, 1434–1449 62 Smith, J. and Bellemare, F. (1987) Effect of lung volume on in vivo contraction characteristics of human diaphragm. J. Appl. Physiol. 62, 1893–1900 63 Druz, W. S. and Sharp, J. T. (1982) Electrical and mechanical activity of the diaphragm accompanying body position in severe chronic obstructive pulmonary disease. Am. Rev. Respir. Dis. 125, 275–280 64 Gross, D., Grassino, A., Ross, W. R. and Macklem, P. T. (1979) Electromyogram pattern of diaphragmatic fatigue. J. Appl. Physiol. 46, 1–7 65 Gandevia, S. C. and McKenzie, D. K. (1986) Human diaphragmatic EMG: changes with lung volume and posture during supramaximal phrenic stimulation. J. Appl. Physiol. 60, 1420–1428 66 Sinderby, C., Lindstrom, L. and Grassion, A. (1995) Automatic assessment of electromyogram quality. J. Appl. Physiol. 79, 1803–1815 67 Sinderby, C., Beck, J., Spahija, J., Weinberg, J. and Grassino, A. (1998) Voluntary activation of the human diaphragm in health and disease. J. Appl. Physiol. 85, 2146–2158 68 Bartolo, A., Roberts, C., Dzwonczyk, R. R. and Goldman, E. (1996) Analysis of diaphragm EMG signals: comparison of gating vs. subtraction for removal of ECG contamination. J. Appl. Physiol. 80, 1898–1902 69 Bolton, C. F. (1993) Clinical neurophysiology of the respiratory system. Muscle Nerve 16, 809–818 70 Beck, J., Sinderby, C., Lindstrom, L. and Grassino, A. (1997) Diaphragm interference pattern EMG and compound muscle action potentials: effects of chest wall configuration. J. Appl. Physiol. 82, 520–530  C

The Authors Journal compilation

 C

2008 Biochemical Society

243

244

Y. M. Luo, J. Moxham and M. I. Polkey

71 Luo, Y. M., Lyall, R. A., Harris, M. L. et al. (2000) Effect of lung volume on oesophageal diaphragm EMG assessed by magnetic phrenic nerve stimulation. Eur. Respir. J. 15, 1033–1038 72 Sinderby, C., Beck, J., Lindstrom, L. H. and Grassino, A. E. (1997) Enhancement of signal quality in esophageal recordings of diaphragm EMG. J. Appl. Physiol. 82, 1370–1377 73 Sharp, J. T., Hammond, M. D., Aranda, A. U. and Rocha, R. D. (1993) Comparison of diaphragm EMG centroid frequencies: esophageal versus chest surface leads. Am. Rev. Respir. Dis. 147, 764–767 74 Grassino, A. E., Whitelaw, W. A. and Milic-Emili, J. (1976) Influence of lung volume and electrode position on electromyography of the diaphragm. J. Appl. Physiol. 40, 971–975 75 American Thoracic Society/European Respiratory Society. (2002) ATS/ERS Statement on respiratory muscle testing. Am. J. Respir. Crit. Care Med. 166, 518–624 76 Lourenco, R. V. and Miranda, J. M. (1968) Drive and performance of the ventilatory apparatus in chronic obstructive lung disease. N. Engl. J. Med. 279, 53–59 77 Juan, G., Calverley, P., Talamo, C., Schnader, J. and Roussos, C. (1984) Effect of carbon dioxide on diaphragmatic function in human beings. N. Engl. J. Med. 310, 874–879 78 Sieck, G. C. and Fournier, M. (1990) Changes in diaphragm motor unit EMG during fatigue. J. Appl. Physiol. 68, 1917–1926 79 Cibella, F., Cuttitta, G., Romano, S., Bellia, V. and Bonsignore, G. (1997) Evaluation of diaphragmatic fatigue in obstructive sleep apnoeas during non-REM sleep. Thorax 52, 731–735 80 Moxham, J., Edwards, R. H. T., Aubier, M. et al. (1982) Changes in EMG power spectrum (high/low ratio) with force fatigue in man. J. Appl. Physiol. 53, 1094–1099 81 Similowski, T., Straus, C., Boistenau, D., Duguet, A. and Derenne, J.-P. (1994) Cervical magnetic stimulation vs transcutaneous electrical phrenic stimulation: comparison of diaphragm action potentials. Am. J. Respir. Crit. Care Med. 149, A129 82 Mills, G. H., Ponte, J., Hamnegard, C. H. et al. (2001) Tracheal tube pressure change during magnetic stimulation of the phrenic nerves as an indicator of diaphragm strength on the intensive care unit. Br. J. Anaesth. 87, 876–884 83 Merton, P. A. and Morton, H. B. (1980) Stimulation of the cerebral cortex in the intact human subject. Nature 1980, 285–227 84 Gea, J., Espadaler, J. M., Guiu, R., Aran, X., Seoane, L. and Broquetas, J. M. (1993) Diaphragmatic activity induced by cortical stimulation: surface versus esophageal electrodes. J. Appl. Physiol. 74, 655–658 85 Sharshar, T., Hopkinson, N. S., Jonville, S. et al. (2004) Demonstration of a second rapidly conducting cortico-diaphragmatic pathway in humans. J. Physiol. 560, 897–908

86 Bigland-Ritchie, B. and Woods, J. J. (1974) Integrated EMG and oxygen uptake during dynamic contractions of human muscles. J. Appl. Physiol. 36, 475–479 87 Bazzy, A. R. and Haddad, G. G. (1984) Diaphragmatic fatigue in unanesthetized adult sheep. J. Appl. Physiol. 57, 182–190 88 Beck, J., Sinderby, C., Lindstrom, L. and Grassino, A. (1998) Effects of lung volume on diaphragm EMG signal strength during voluntary contractions. J. Appl. Physiol. 85, 1123–1134 89 Aubier, M., Trippenbach, T. and Roussos, C. (1981) Respiratory muscle fatigue during cardiogenic shock. J. Appl. Physiol. 51, 499–508 90 Roussos, C. S. and Macklem, P. T. (1977) Diaphragmatic fatigue in man. J. Appl. Physiol. 43, 189–197 91 Dayer, M. J., Hopkinson, N. S., Ross, E. T. et al. (2006) Does symptom-limited cycle exercise cause low frequency diaphragm fatigue in patients with heart failure? Eur. J. Heart Failure 8, 68–73 92 Laghi, F., Cattapan, S. E., Jubran, A. et al. (2003) Is weaning failure caused by low-frequency fatigue of the diaphragm?Am. J. Respir. Crit. Care Med. 167, 120–127 93 Polkey, M. I., Kyroussis, D., Hamnegard, C.-H. et al. (1997) Diaphragm performance during maximal voluntary ventilation in chronic obstructive pulmonary disease. Am. J. Respir. Crit. Care Med. 155, 642–648 94 Polkey, M. I., Kyroussis, D., Keilty, S. E. J. et al. (1995) Exhaustive treadmill exercise does not reduce twitch transdiaphragmatic pressure in patients with COPD. Am. J. Respir. Crit. Care Med. 152, 959–964 95 De Jonghe, B., Sharshar, T., Lefaucheur, J. P. et al. (2002) Paresis acquired in the intensive care unit: a prospective multicenter study. JAMA, J. Am. Med. Assoc. 288, 2859–2867 96 Polkey, M. I. and Moxham, J. (2001) Clinical aspects of respiratory muscle dysfunction in the critically ill. Chest 119, 926–939 97 Vassilakopoulos, T. and Petrof, B. J. (2004) Ventilatorinduced diaphragmatic dysfunction. Am. J. Respir. Crit. Care Med. 169, 336–341 98 Vincken, W., Guilleminault, C., Silvestri, L., Cosio, M. and Grassino, A. (1987) Inspiratory muscle activity as a trigger causing the airways to open in obstructive sleep apnea. Am. Rev. Respir. Dis. 135, 372–377 99 Luo, Y. M., Wu, H. D., Tang, J. et al. (2008) Neural respiratory drive during apnoeic events in obstructive sleep apnoea. Eur. Respir. J. 31, 650–657 100 Tang, J., Jolley, C., Wu, H. D. et al. (2007) Distinguishing central sleep apnea from obstructive sleep apnea with esophageal diaphragm EMG. Am. J. Respir. Crit. Care Med. 175, A579 101 Steier, J., Jolley, C., Wu, H. D. et al. (2007) Neural respiratory drive during airway occlusion in obstructive sleep apnea (OSA). Am. J. Respir. Crit. Care Med. 175, A69

Received 2 October 2007/10 January 2008; accepted 14 January 2008 Published on the Internet 12 September 2008, doi:10.1042/CS20070348