Enhancement of Frequency Stability Using

0 downloads 0 Views 4MB Size Report
Oct 13, 2016 - ... 24 V at the system anchor and using Keysight N9030A signal analyzer .... its modal frequency) acquired with the oscilloscope Agilent DSC-X ...
sensors Article

Enhancement of Frequency Stability Using Synchronization of a Cantilever Array for MEMS-Based Sensors Francesc Torres *, Arantxa Uranga, Martí Riverola, Guillermo Sobreviela and Núria Barniol Electrical Engineering Department, Universitat Autònoma de Barcelona, Edifici Q, Campus UAB Bellaterra, Cerdanyola del Vallès 08193, Spain; [email protected] (A.U.); [email protected] (M.R.); [email protected] (G.S.); [email protected] (N.B.) * Correspondence: [email protected]; Tel.: +34-93-581-8463 Academic Editor: Stefano Mariani Received: 6 July 2016; Accepted: 8 October 2016; Published: 13 October 2016

Abstract: Micro and nano electromechanical resonators have been widely used as single or multiple-mass detection sensors. Smaller devices with higher resonance frequencies and lower masses offer higher mass responsivities but suffer from lower frequency stability. Synchronization phenomena in multiple MEMS resonators have become an important issue because they allow frequency stability improvement, thereby preserving mass responsivity. The authors present an array of five cantilevers (CMOS-MEMS system) that are forced to vibrate synchronously to enhance their frequency stability. The frequency stability has been determined in closed-loop configuration for long periods of time by calculating the Allan deviation. An Allan deviation of 0.013 ppm (@ 1 s averaging time) for a 1 MHz cantilever array MEMS system was obtained at the synchronized mode, which represents a 23-fold improvement in comparison with the non-synchronized operation mode (0.3 ppm). Keywords: MEMS; synchronization; resonators; CMOS-MEMS; cantilevers; arrays; coupling

1. Introduction Nano and micro electromechanical systems (M/NEMS) have been widely used for single or multiple mass detection due to their small effective masses and high resonance frequencies. Ultimate limits in mass sensing have been reported with NEMS-based resonators [1]. In dynamic or resonant mode, one of the main characteristics that limits the mass resolution is the stability of the resonance frequency of the M/NEMS resonator. This limitation comes from multiple sources of noise, which become more important with the size reduction to the nanoscale of the NEMS devices as has been recently reported [2]. Synchronization is described as an adjustment of rhythms of oscillating objects due to their weak interaction [3]. From the first work of Huygens [4] until the present day, there has been great effort to provide a comprehensive theoretical basis to synchronization [5] and, derived from studies involving collective systems, mathematical techniques to extract from the data the pure synchronous behavior from heterodox phenomena [6]. Recently, synchronization phenomena have become an interesting issue in a wide range of scientific fields [7–9]. In the field of M/NEMS, synchronization phenomena are rising in importance and a lot of scientific works have appeared using a few synchronized elements [10], a lot of them [8], or using synchronization to enhance the mass detection performance of MEMS-based sensors [11]. Special efforts have focused on studying synchronization based on different methods of interaction between MEMS and its influence on frequency stability [12]. There are important works related to this phenomenon dealing with mechanical interactions [11] or electrical interaction between MEMS [13,14], Sensors 2016, 16, 1690; doi:10.3390/s16101690

www.mdpi.com/journal/sensors

Sensors 2016, 16, 1690

2 of 15

new sensing strategies like localized modes [15], or stiffness change [16]. In addition, there is a parallel effort to enhance the sensors’ performance, trying to use non-linear phenomena to overcome noise limitations [17–19]. We have focused our study on trying to take advantage of a synchronized MEMS system in a closed-loop oscillator to enhance the frequency stability over time using a fully integrated CMOS system, with electrical and mechanical interaction between resonating elements. An example of the utility of synchronization is that it allows for obtaining a higher performance single or multiple mass sensor. This paper is organized as follows: Section 2 describes the design of the system (based on five cantilevers that are mechanically connected), the fabrication method (based on CMOS technology), and the electrical characterization of the system. In this Section 2, the modal frequencies of the five cantilevers (obtained in an open-loop configuration), as well as the frequency at which the system vibrates at closed-loop configuration with its stability in time, are provided. From these results the mass sensitivity of the system (in a closed-loop configuration) is computed. Section 3 is devoted to characterizing the stability of the frequency in time in a closed-loop configuration when the system is forced to synchronously vibrate using an external force. The authors present two different ways to force the synchronization, the first one acting over one individual cantilever at its modal frequency and the other one acting over the same cantilever but at the frequency at which the system vibrates in a closed-loop configuration. In this section, the authors also present results of the frequency stability over time and the lower limit of mass resolution. Section 4 is devoted to the synchronization process itself, computing the Arnold tongue for both synchronization methods and including a discussion of the mass sensing limitations working synchronously. Finally, Section 4 also presents a short discussion of the possible implications of the synchronization phenomena with noise. 2. Materials and Methods 2.1. CMOS-MEMS System Design and Fabrication The CMOS-MEMS system is fully integrated using Austria MicroSystems (AMS) 0.35 µm CMOS technology [20]. The whole system is shown in Figure 1, which comprises five cantilevers electrostatically sensed and actuated out of plane and a CMOS transimpedance amplification circuitry [21]. The cantilevers were fabricated using the Metal 4 layer of AMS technology, clamped together, and mechanically connected through an overhang (see Figure 1a). Each cantilever is 26 µm long, 1.45 µm wide and 0.925 µm thick; the overhang is 6 µm wide and has the same thickness as the cantilevers. Read-out drivers were fabricated using the Metal 3 layer of AMS technology, 1 µm below Metal 4 layer, which will be the gap distance between cantilevers and read-out or actuation drivers. There is one common read-out driver (CD) connected to the transimpedance amplifier to sense the five cantilevers at the same time. This common driver is placed at the tip of the cantilevers (see Figure 1c). There is also one individual driver (ID) for each cantilever, connected directly to our characterization setup without amplification, which allows the individual actuation/sensing. There is a surrounding shield for all drivers connected to the ground, which is made of a Metal 3 layer, in order to minimize the external noise (frame wrapping common and individual drivers, see Figure 1c). The definition of the MEMS cantilevers is completely done in the AMS foundry. An in-house releasing process of the MEMS movable parts is done on the received bare chips, which require only one post-processing step in order to eliminate the sacrificial oxide surrounding the structural metal parts of the MEMS resonator (in our case, the five cantilevers). This post-processing step consists of immersing the chip in buffered hydrofluoric acid (BHF) and then rinsing it with deionized water and isopropyl alcohol to avoid stiction between the cantilevers. The engraving time has to be carefully calibrated to avoid undesirable under-etching, taking into account that the deeper the layer, the longer the engraving time. The chips are protected by the CMOS passivation layer and, since a window is opened in the passivation layer over the multicantilever system, obtaining in this way a

Sensors 2016, 16, 1690

3 of 15

pool-like structure, BHF engraves only the chip zones where the multicantilever system is, releasing theSensors MEMS system. 2016, 16, 1690 3 of 15

(a)

(b)

(c)

Figure 1. (a) Optical image of the five released cantilevers (26 μm long, 1.45 μm wide, 0.925 μm thick) Figure 1. (a) Optical image of the five released cantilevers (26 µm long, 1.45 µm wide, 0.925 µm thick) in which the overhang can be seen (6 μm wide, 0.925 μm thick). The five cantilevers are numbered to in which the overhang can be seen (6 µm wide, 0.925 µm thick). The five cantilevers are numbered to reference them in the main text; (b) Optical image of the whole system, including the cantilever array reference them in the main text; (b) Optical image of the whole system, including the cantilever array and CMOS transimpedance amplifier; (c) Image of the layout of the drivers with the individual and CMOS transimpedance amplifier; (c) Image of the layout of the drivers with the individual drivers drivers (ID) and common driver (CD) and their access. CD common driver is below the tip of the (ID) and common driver (CD) and their access. CD common driver is below the tip of the cantilevers cantilevers and the ID individual drivers are between the middle and near-to-anchor part of the and the ID individual drivers are between the middle and near-to-anchor part of the cantilevers. cantilevers.

The releasingpost-process post-processstep stepand and the the dimension dimension tolerances used The releasing tolerancesfrom fromthe theCMOS CMOStechnology technology used add somesort sortofofanisotropy anisotropytoto the the modal modal frequencies frequencies of add some of the the cantilevers cantilevers(providing (providinga adifferent different resonantfrequency frequencyfor for each each of the fact, the the materials around the system and the resonant the cantilevers). cantilevers).In In fact, materials around the system and pool-like structure break the symmetric surrounding of the five cantilevers under the BHF releasing the pool-like structure break the symmetric surrounding of the five cantilevers under the BHF process,process, varyingvarying the dimensions of each cantilever in relation to its and, therefore, releasing the dimensions of each cantilever in relation to neighbors its neighbors and, therefore, producing a dispersion of modal frequencies. This inconvenience could be overcome by tuning thethe producing a dispersion of modal frequencies. This inconvenience could be overcome by tuning resonance frequency by applying DC voltage at the individual drivers, taking advantage of the resonance frequency by applying DC voltage at the individual drivers, taking advantage of the spring spring softening effect. softening effect. CMOS-MEMSSystem SystemElectrical ElectricalCharacterization Characterization 2.2.2.2. CMOS-MEMS measures have been performed in vacuum conditions. To characterize the natural AllAllthethemeasures have been performed in vacuum conditions. To characterize the natural resonance frequency of each cantilever, we have measured the thermomechanical noise at the resonance frequency of each cantilever, we have measured the thermomechanical noise at the common common driver, applying a DC voltage of 24 V at the system anchor and using Keysight N9030A driver, applying a DC voltage of 24 V at the system anchor and using Keysight N9030A signal analyzer signal analyzer with 10 Hz of IF bandwidth. Five different frequency peaks corresponding to the five with 10 Hz of IF bandwidth. Five different frequency peaks corresponding to the five cantilever cantilever MEMS resonators have been obtained (see Figure 2) and an additional one is due to MEMS resonators have been obtained (see Figure 2) and an additional one is due to external noise. external noise. Applying a DC voltage at each individual driver (see Figure 1), we can use the spring Applying a DC voltage at each (see Figure 1), we can use number the spring softening softening effect to relate eachindividual peak to thedriver corresponding cantilever. Peak 5 (Figure 2) iseffect a to sum relateofeach peak to the corresponding cantilever. Peak number 5 (Figure 2) is a sum of the peak the peak corresponding to cantilever number five and the parasitic external noise corresponding to by cantilever number five and the parasitic noise (demonstrated measuring (demonstrated measuring the thermomechanical noise external at 0 V of DC effective voltage), by resulting in thea thermomechanical noise at 0 V of DC effective voltage), resulting in a higher magnitude peak higher magnitude peak compared with the others. compared with the In order to others. characterize the frequency stability of the system, we performed closed-loop In order totaking characterize the frequency stability ofamplifier the system, performed measurement, the signal from the transimpedance (TIA)we connected to theclosed-loop common measurement, taking the signal from the transimpedance (TIA) connected to the common driver, adding DC voltage of 24 V through a bias-tee, andamplifier driving this signal to the anchor of the system (see Figure 3a). In this configuration, the system is self-oscillating (see Figure 3b). The driver, adding DC voltage of 24 V through a bias-tee, and driving this signal to the anchor of the system frequency was In acquired using a Hewlett frequency counter (Keisight (see Figure 3a). this configuration, thePackard system53131A is self-oscillating (see Figure 3b). Technologies, The frequency Santa Rosa, using CA, USA). was acquired a Hewlett Packard 53131A frequency counter (Keisight Technologies, Santa Rosa, We can see in Figure 4a the stability of the frequency of oscillation, taking measurements for CA, USA). two (each measure is taken every 0.1frequency s), in which we can appreciate a long time for drift Wehours can see in Figure 4a the stability of the of oscillation, taking measurements two (corresponding to a linear drift of 0.1 Hz/s). The system auto-oscillates at the frequency hours (each measure is taken every 0.1 s), in which we can appreciate a long time drift (corresponding to 0.1 peak number of Figure 2. The corresponding Allan corresponding Deviation from to time to corresponding a linear drift of Hz/s). The3system auto-oscillates at the frequency peak measurements every 0.1 s, shown in Figure 4b, has a value below 1 ppm at 1 s of averaging time. number 3 of Figure 2. The corresponding Allan Deviation from time measurements every 0.1 s, shown in Figure 4b, has a value below 1 ppm at 1 s of averaging time.

Sensors 2016, 16, 1690

Sensors 2016, 16, 1690 Sensors 2016, 16, 1690 Sensors 2016, 16, 1690

4 of 15

4 of 15 4 of 15 4 of 15

Figure 2. Thermomechanical noisemeasured measured at the applying a DC voltage of 24 of V at Figure 2. Thermomechanical noise thecommon commondriver driver applying a DC voltage 24 V Figure 2. Thermomechanical noise measured at the common driver applying a DC voltage of 24 V at the system anchor.We Weflag flagthe the five five peaks to identify each one with itsitscorresponding cantilever at the system anchor. peaks identify each one with corresponding cantilever Figure 2. Thermomechanical noise measured at the common driver applying a DC voltage of 24 V at the system anchor. We flag the five peaks to identify each one with its corresponding cantilever (shown in Figure 1a). Peak number 5peaks is enhanced by the parasitic noise peak. Inset: parasitic noise (shown in Figure 1a).1a). Peak number 5 5is by parasitic noise peak. Inset: parasitic noise the system anchor. We flag the five to identify each one with itspeak. corresponding cantilever (shown in Figure Peak number isenhanced enhanced by the the parasitic noise Inset: parasitic noise peak measured at 0ofVDC of DC voltage.5 is enhanced by the parasitic noise peak. Inset: parasitic noise (shown in Figure 1a). Peak number peakpeak measured at 0 V voltage. measured at 0 V of DC voltage. peak measured at 0 V of DC voltage.

(a) (b) (a) (b) (a) of the closed-loop setup, showing the capability (b) of additional stimuli at Figure 3. (a) Schematic Figure 3. (a) Schematic of the closed-loop setup, showing the capability of additional stimuli at Figure 3. (a) Schematic of the closed-loop setup, showing the capability of additional stimuli individual drivers and of an the optical image of the showing five-cantilever array; (b) Time response of the Figure 3. (a) Schematic closed-loop the capability additional stimuli at at individual drivers and an optical image ofsetup, the five-cantilever array; (b)ofTime response of the individual drivers and an optical image ofvoltage the of five-cantilever array; (b) Time of the closed-loop closed-loop oscillation (applying a DC of 24five-cantilever V). individual drivers and(applying an optical image the array; (b)response Time response of the closed-loop oscillation a DC voltage of 24 V). oscillation (applying a DC(applying voltage of 24 V). closed-loop oscillation a DC voltage of 24 V).

(a) (b) (a) (b) (a) (b) Figure 4. (a) Frequency stability measured in closed-loop configuration for 120 min. There is a clear

Figure 4. (a) Frequency stability measured in closed-loop configuration for 120 min. There is a clear long term drift; (b) Allan deviation in ppminfor the closed-loop measurements with Vdc = is 24 Vclear and Figure 4. (a) Frequency stability measured closed-loop configuration for 120with min.Vdc There long drift; (b) Allan deviation in ppmin for the closed-loop measurements = 24 aVis and Figure 4.term (a) Frequency stability measured closed-loop configuration for 120 min. There a clear taking measures every 0.1 s. The minimum Allan deviation is near 1 s of averaging time. long term drift; (b) Allan deviation in ppm for the closed-loop measurements withtime. Vdc = 24 V and s. The minimum deviation is near 1 s of averaging longtaking term measures drift; (b) every Allan0.1 deviation in ppm Allan for the closed-loop measurements with Vdc = 24 V and taking measures every 0.1 s. The minimum Allan deviation is near 1 s of averaging time.

taking measures every 0.1 s. The minimum Allan deviation is near 1 s of averaging time.

Sensors 2016, 16, 1690

5 of 15

2.3. CMOS-MEMS System Mass Resolution Following the same procedure as [21], we can calculate the mass sensitivity using the standard Equation (1), where meff is the effective mass of the system, f 0 is the resonance frequency in auto-oscillation mode, and ∆f is the dispersion of frequencies, ∆f = f 0 ·σ, where σ is the Allan Deviation for the specific averaging time. For our system under self-oscillation conditions, f 0 = 1.102 MHz, meff = 1.1 × 10−10 kg and the frequency dispersion is ∆f = 0.3 Hz, which corresponds to σ = 3.5 × 10−7 (at an averaging time of 1 s). Consequently, the minimum achievable mass detection, or mass resolution, is 60 ag (attograms). 2me f f ∆m = ∆f (1) f0 3. Characterization of the Cantilever Array under Synchronization As is reported in the works mentioned above (see [13], for example), one of the special features of the synchronized state is the enhanced frequency stability. In our CMOS-MEMS system, the frequency dispersion between the individual cantilevers due to the post-process and dimensions tolerance of the fabrication is too high to allow for natural synchronization. For natural synchronization it is mandatory to have a commensurable relationship n:m, where n and m are integers, between resonance frequencies of the oscillators to be synchronized [3]. Measuring the closed-loop oscillation with our five-cantilever array CMOS MEMS system, we are not able to observe synchronization. Despite the fact that the system oscillates (at closed-loop configuration) at one individual frequency (and not five), the fluctuations of it and the long time drift are high enough to discard a synchronized state. Due to this inconvenience, we changed our measurement procedure. We used the closed-loop configuration for the following measurements, in order to have a self-sustained system (a system that oscillates taking energy from a source, in our case, a DC voltage) [3]. We achieved synchronization by applying an external electrical force (stimulation) through a sinusoidal signal (from the function generator, Agilent 81150A, (Keisight Technologies, Santa Rosa, CA, USA) to an individual driver (ID, see Figure 1). We used two of the five cantilevers, numbers 2 and 4 (see Figure 1a), as candidates for this external forced synchronization. We used two methods; one is based on applying an external force to one of the individual drivers using the same frequency as the modal frequency of the corresponding individual cantilever. In this case the external force induces the cantilevers’ synchronization (see Section 3.1). The other method is based on applying this force, also on an individual driver, but with the same frequency as the self-oscillation state has. In this case the external force synchronizes with our self-sustained system (see Section 3.2). A discussion of these two methods of external excitation is presented in Section 4. 3.1. Using an External Force Applied to One of the Individual Cantilevers, at the Same Frequency as the Modal Frequency of the Corresponding Cantilever The external force is applied (as we said before) on one individual driver (ID) at the same frequency as the corresponding cantilever modal frequency with different amplitudes. The time-dependent signal of the system response is acquired, in a closed-loop configuration, at the common driver, for an applied DC voltage of 18 V and for different stimulation amplitudes. In Figure 5a we can see the time domain response for different stimulation amplitudes at the cantilever number 4 (and at its modal frequency) acquired with the oscilloscope Agilent DSC-X 3054A. We have a sum of two signals with different frequencies resulting in a pulsed time-dependent signal, until the amplitude reaches 500 mV (peak-to-peak), where the signal stabilizes at a single frequency, which corresponds to the modal frequency of the stimulated cantilever (number 4). If we analyze the signal (using Fourier transform) at 500 mV of excitation amplitude, we obtain a single peak at the modal frequency of the stimulated cantilever, which is a signature of synchronized state. On the other hand, if we analyze the pulsed signal (using Fourier transform) we conclude that the pulse is due to the sum of two signals, one at 1.076 MHz, corresponding to the modal frequency of cantilever number 4

Sensors 2016, 16, 1690 Sensors 2016, 16, 1690

6 of 15 6 of 15

due to the sum of two signals, one at 1.076 MHz, corresponding to the modal frequency of cantilever number 4 of (frequency of stimulation), and another signal with a frequency of 1.108 MHz, (frequency stimulation), and another signal with a frequency of 1.108 MHz, corresponding to corresponding to frequency the self-oscillation frequency ofadditional the system without additional (both the self-oscillation of the system without stimuli (both applying stimuli a DC voltage applying a DC on voltage 18 V). Based on measurements, we decided to mV use aofstimulation of of 18 V). Based theseof measurements, wethese decided to use a stimulation of 650 amplitude to 650 mV of amplitude to ensure, as much as possible, a single value of the frequency for the read-out ensure, as much as possible, a single value of the frequency for the read-out signal. Figure 5b, in an signal. Figure 5b, in an open-loop configuration, shows the effectnoise on thermomechanical when open-loop configuration, shows the effect on thermomechanical when applying annoise excitation applying an excitation to the corresponding cantilever (row A corresponds to the pure to the corresponding cantilever (row A corresponds to the pure thermomechanical noise, without thermomechanical noise, without excitation; row B corresponds to the thermomechanical noise plus excitation; row B corresponds to the thermomechanical noise plus excitation of cantilever number 4 excitation of cantilever number 4 and rows C and D to cantilever number 2). and rows C and D to cantilever number 2).

(a)

(b)

Figure 5. (a) Time response signal of the closed-loop self-oscillation of the system with stimulation Figure 5. (a) Time response signal of the closed-loop self-oscillation of the system with stimulation on on cantilever number 4 with voltage V and using different amplitudes peak-to-peak: A,mV; 350 cantilever number 4 with DC DC voltage of 18ofV18 and using different amplitudes peak-to-peak: A, 350 mV; 450C, mV; 500D, mV; 600(b) mV; (b) Thermomechanical noise plus excitation with DC voltage B, 450B,mV; 500C,mV; 600D, mV; Thermomechanical noise plus excitation with DC voltage of 24 V of 24 V in open-loop configuration: A, without excitation; B, excitation on cantilever number 4 at its in open-loop configuration: A, without excitation; B, excitation on cantilever number 4 at its modal modal frequency with 650 mV of amplitude, where we can see that peak number 4 stands above the frequency with 650 mV of amplitude, where we can see that peak number 4 stands above the others; others; C, excitation on cantilever number 2 at its modal frequency with 650 mV of amplitude, where C, excitation on cantilever number 2 at its modal frequency with 650 mV of amplitude, where we can we that can see that peak number standsthe above the D, others; D, excitation on cantilever at the see peak number 2 stands2above others; excitation on cantilever numbernumber 2 at the2modal modal frequency of cantilever number 3, where we can see that the five peaks remain and a large and frequency of cantilever number 3, where we can see that the five peaks remain and a large and sharp sharp peak appears as number 3. peak appears as number 3.

Figure 5b shows that, when we apply an excitation over an individual cantilever at its modal Figure 5b shows that, when we apply an excitation over an individual cantilever at its modal frequency (Figure 5b, rows B and C), the peaks corresponding to the other cantilevers disappear and frequency (Figure 5b, rows B and C), the peaks corresponding to the other cantilevers disappear and only the one corresponding to the excited cantilever remains; moreover, this peak has higher power only the one corresponding to the excited cantilever remains; moreover, this peak has higher power than in the case without excitation (for example, we can compare peak number 2 of rows A and C or than in the case without excitation (for example, we can compare peak number 2 of rows A and C or peak number 4 in rows A and B of Figure 5b). We deduce that, in this case, the system is peak number 4 in rows A and B of Figure 5b). We deduce that, in this case, the system is synchronized synchronized and oscillates at the modal frequency of the excited cantilever. When the excitation is and oscillates at the modal frequency of the excited cantilever. When the excitation is applied to applied to the cantilever but at a different frequency, the five peaks remain and no traces of the cantilever but at a different frequency, the five peaks remain and no traces of synchronization synchronization appear. An example of this is shown in row D on Figure 5b, where excitation is appear. An example of this is shown in row D on Figure 5b, where excitation is applied over cantilever applied over cantilever number 2 but at the frequency corresponding to cantilever 3; we can see the number 2 but at the frequency corresponding to cantilever 3; we can see the five individual peaks as five individual peaks as in row A of Figure 5b (except a sharper peak coming from the excitation in row A of Figure 5b (except a sharper peak coming from the excitation itself), contrary to row C on itself), contrary to row C on Figure 5b, in which only peak number 2 remains. Figure 5b, in which only peak number 2 remains. Following the same steps mentioned in Section 2 but now using stimulation, we measured the Following the same steps mentioned in Section 2 but now using stimulation, we measured the frequency stability of the system for 120 min. The procedure was as follows: we stimulated one frequency stability of the system for 120 min. The procedure was as follows: we stimulated one cantilever, number 2 or number 4, applying the signal at its own individual driver and at its own cantilever, number 2 or number 4, applying the signal at its own individual driver and at its own modal frequency; the closed-loop was done as previously, taking the signal from the modal frequency; the closed-loop was done as previously, taking the signal from the transimpedance transimpedance amplifier connected to the common driver and driving this signal to the anchor of amplifier connected to the common driver and driving this signal to the anchor of the system adding, the system adding, through a bias-tee, a DC voltage of 24 V. We observed that, when the stimulation through a bias-tee, a DC voltage of 24 V. We observed that, when the stimulation is at the modal is at the modal frequency of the corresponding stimulated cantilever, the whole system oscillates at frequency of the corresponding stimulated cantilever, the whole system oscillates at this frequency; this frequency; otherwise the system oscillates at the frequency of cantilever number three (as otherwise the system oscillates at the frequency of cantilever number three (as without stimulation), without stimulation), as we can see in Figure 6 for both cantilevers. To discard the possible effect of

Sensors 2016, 16, 1690

7 of 15

Sensors Sensors2016, 2016,16, 16,1690 1690

77of of15 15

as we can see in Figure 6 for both cantilevers. To discard the possible effect of the stimulus over the the over the driver, could produce aa read-out signal directly the the stimulus stimulus over the common common driver,a which which could produce read-out signal generator directly from from the common driver, which could produce read-out signal directly from the signal instead signal generator instead of the cantilevers system, we performed stimulus at the modal frequency signal generator instead of the cantilevers system, we performed stimulus at the modal frequency of the cantilevers system, we performed stimulus at the modal frequency and at two frequencies and two frequencies near the modal, above and below it. when the is and at at two frequencies nearand theanother modal, one one above andanother another below it. Only Only when the stimulus stimulus is at at near the modal, one above below it. Only when the stimulus is at the modal frequency the modal frequency does the read-out signal have this frequency; otherwise, it has the frequency of the modal frequency doeshave the read-out signal have this frequency; otherwise, it has of does the read-out signal this frequency; otherwise, it has the frequency of the the frequency closed-loop the closed-loop without stimulus. Added to this fact, there is another and more important effect: the the closed-loop without stimulus. Added to this fact, there is another and more important effect: the without stimulus. Added to this fact, there is another and more important effect: the high frequency high stabilization the stimulus is frequency of high frequency frequency stabilization obtained when the stimulus is at atofthe the modal frequency of the the stabilization obtained when theobtained stimuluswhen is at the modal frequency themodal corresponding cantilever. corresponding cantilever. For instance, from Figure 6a, the time frequency stability corresponding to corresponding cantilever. For instance, from Figure 6a, the time frequency stability corresponding to For instance, from Figure 6a, the time frequency stability corresponding to case A is around 0.05 Hz/s; −7 −7 Hz/s. case A is around 0.05 Hz/s; for case B it is 0.005 Hz/s and for case C it is only 5 × 10 In Figure 7, − 7 case A is around 0.05 Hz/s; for case B it is 0.005 Hz/s and for case C it is only 5 × 10 Hz/s. In Figure 7, for case B it is 0.005 Hz/s and for case C it is only 5 × 10 Hz/s. In Figure 7, we have adjusted the we have adjusted the graphics of Figure 6 to highlight this effect, scaling the frequencies to the we have of adjusted Figure to highlight this effect,toscaling the frequencies to the graphics Figure 6the to graphics highlight of this effect,6 scaling the frequencies the frequency of the starting frequency of the starting point. When the stimulus is at a different frequency than the modal, the frequency of the starting point. When the stimulus is at a different frequency than the modal, the point. When the stimulus is at a different frequency than the modal, the frequency stabilization does frequency stabilization does not occur. frequency stabilization does not occur. not occur.

(a) (a)

(b) (b)

Figure frequencies for stimulus at number 22 (modal frequency, ff0202 == Figure 6. 6. (a) (a) Closed-loop frequencies forfor stimulus at cantilever cantilever number (modal frequency, Figure 6. (a)Closed-loop Closed-loop frequencies stimulus at cantilever number 2 (modal frequency, 1.0549 MHz) with different frequencies: A, at 1.06 MHz (above f 02 ); B, at 1.04 MHz (below f 02 ); C, at 1.0549 MHz) with different frequencies: A, at 1.06 MHz (above f 02 ); B, at 1.04 MHz (below f 02 ); C, at f 02 = 1.0549 MHz) with different frequencies: A, at 1.06 MHz (above f 02 ); B, at 1.04 MHz (below 02 ); (b) Closed-loop frequencies for stimulus at cantilever number 4 (modal 1.0549 MHz (exactly at f 02 ); (b) Closed-loop frequencies for stimulus at cantilever number 4 (modal 1.0549 MHz (exactly at f f 02 ); C, at 1.0549 MHz (exactly at f 02 ); (b) Closed-loop frequencies for stimulus at cantilever number 4 with different A, ff0404);); B, at frequency, ff0404 == 1.07348 1.07348 MHz) with different frequencies: A, at at 1.08 1.08 MHz MHz (above B, at 1.06 1.06 MHz frequency, (modal frequency, f 04 =MHz) 1.07348 MHz) withfrequencies: different frequencies: A, at(above 1.08 MHz (above f 04MHz ); B, (below ff0404);); C, at (exactly at (below C, at 1.07348 1.07348 MHz (exactlyMHz atff0404).). at 1.06 MHz (below f ); MHz C, at 1.07348 (exactly at f ). 04

(a) (a)

04

(b) (b)

Figure Figure 7. 7. (a) (a) Closed-loop Closed-loop frequency frequency scaled scaled at at the the starting starting point point for for the the case case of of stimulus stimulus at at cantilever cantilever Figure 7. (a) Closed-loop frequency scaled at the starting point for the case of stimulus at cantilever number 2 with different frequencies: A, at 1.06 MHz; B, at 1.04 MHz; C, at 1.0549 number 2 with different frequencies: A, at 1.06 MHz; B, at 1.04 MHz; C, at 1.0549 MHz; MHz; (b) (b) number 2 with different frequencies: A, at 1.06 MHz; B, at 1.04 MHz; C, at 1.0549 MHz; (b) Closed-loop Closed-loop frequency scaled at the starting point for the case of stimulus at cantilever number 4 Closed-loop frequency scaled at the starting point for the case of stimulus at cantilever number frequency scaled at the starting point for the case of stimulus at cantilever number 4 with different4 with different frequencies: A, at 1.08 MHz; B, at 1.06 MHz; C, at 1.07348 MHz. with different frequencies: A, at 1.08 MHz; B, at 1.06 MHz; C, at 1.07348 MHz. frequencies: A, at 1.08 MHz; B, at 1.06 MHz; C, at 1.07348 MHz.

Once Once we we measured measured the the closed-loop closed-loop frequency frequency with with an an applied applied stimulus, stimulus, the the next next step step was was to to Once we measured the closed-loop frequency with an applied stimulus,every the next step wasthe to evaluate the frequency stability calculating the Allan deviation (with measures 0.1 s) under evaluate the frequency stability calculating the Allan deviation (with measures every 0.1 s) under the evaluate the frequencycan stability calculating the AllanAdeviation (with measuresbehavior every 0.1 s)that under the same same conditions. conditions. We We can see see in in Figure Figure 88 that that cases cases A and and BB present present similar similar behavior to to that found found without synchronization (Figure 4b), and that case C improves the Allan deviation by almost without synchronization (Figure 4b), and that case C improves the Allan deviation by almost two two orders orders of of magnitude magnitude (@ (@ 11 ss of of averaging averaging time). time). Due Due to to the the fact fact that that the the stability stability of of the the frequency frequency is is

Sensors 2016, 16, 1690

8 of 15

same conditions. We can see in Figure 8 that cases A and B present similar behavior to that found without synchronization (Figure 4b), and that case C improves the Allan deviation by almost8 two Sensors 2016, 16, 1690 of 15 orders of magnitude (@ 1 s of averaging time). Due to the fact that the stability of the frequency is very very good for atime, long i.e., time, i.e.,isthere is no frequency C in6Figures 6 and 7), the Allan good for a long there no frequency drift (seedrift case(see C incase Figures and 7), the Allan deviation deviation decreases continuously in our measurement timewe window, aretonot able tothe observe decreases continuously and, in ourand, measurement time window, are notwe able observe Allan the Allan limit deviation due tonoise the Flicker floor). addition to that and further deviation due tolimit the Flicker (Flickernoise floor).(Flicker In addition to In that and further supporting our supporting our isclaim that effect there between is no direct effectand between stimulus and read-out, Allan claim that there no direct stimulus read-out, the Allan deviation forthe stimulus deviation stimulusisatat different frequencies is at the same of magnitude asstimulus for the at differentfor frequencies the same order of magnitude as for the order self-oscillation without self-oscillation without stimulus (see Figure 8). (see Figure 8).

(a)

(b)

Figure 8. 8. (a) every 0.1 0.1 ss Figure (a) Allan Allan deviation deviation in in ppm ppm stimulating stimulating the the cantilever cantilever number number 2, 2, taking taking measures measures every and for stimulus frequencies of: A, at 1.06 MHz; B, at 1.04 MHz; C, at 1.0549 MHz; (b) Allan deviation and for stimulus frequencies of: A, at 1.06 MHz; B, at 1.04 MHz; C, at 1.0549 MHz; (b) Allan deviation in ppm ppm stimulating the cantilever cantilever number number 4, 4, taking taking measures measures every every 0.1 0.1 ss and in stimulating the and for for stimulus stimulus frequencies frequencies of: A, at 1.06 MHz; B, at 1.08 MHz; C, at 1.07348 MHz. of: A, at 1.06 MHz; B, at 1.08 MHz; C, at 1.07348 MHz.

From Figure 8a, the frequency stability at an averaging time of 1 s, for the case of the stimulus From Figure 8a, the frequency stability at an averaging time of 1 s, for the case of the stimulus applied at cantilever number 2 (at its modal frequency), is ∆f = 0.016 Hz (σ = 1.5 × 10−−88), more or less applied at cantilever number 2 (at its modal frequency), is ∆f = 0.016 Hz (σ = 1.5 × 10 ), more or less the same value as for the stimulus applied at cantilever number 4. Using Equation (1), for the the same value as for the stimulus applied at cantilever number 4. Using Equation (1), for the stimulus stimulus at cantilever number 2 or 4, we achieve a minimum detectable mass of 3 ag. We can further at cantilever number 2 or 4, we achieve a minimum detectable mass of 3 ag. We can further decrease decrease the final mass resolution using a higher averaging time. For instance, with an averaging the final mass resolution using a higher averaging time. For instance, with an averaging time of 100 s, time of 100 s, the frequency dispersion is ∆f = 0.0011 Hz, and consequently the minimum achievable the frequency dispersion is ∆f = 0.0011 Hz, and consequently the minimum achievable mass is 0.2 ag. mass is 0.2 ag. In summary, from this synchronization technique, using forced cantilevers with In summary, from this synchronization technique, using forced cantilevers with stimulation at their stimulation at their own modal frequency, we can achieve a minimum detectable mass 300 times own modal frequency, we can achieve a minimum detectable mass 300 times lower than in the case lower than in the case without stimulus with a long averaging time. without stimulus with a long averaging time. These minimum achievable masses are calculated using Equation (1). Due to the fact that we These minimum achievable masses are calculated using Equation (1). Due to the fact that we have have a synchronized system through the presence of an external force, these minimum achievable a synchronized system through the presence of an external force, these minimum achievable masses masses have to be taken as a lower limit of achievable masses. See the discussion of this fact in have to be taken as a lower limit of achievable masses. See the discussion of this fact in Section 4.2. Section 4.2. 3.2. Using an External Force Applied to One of the Individual Cantilevers at the Self-Oscillating Frequency of 3.2.System Using an External Force Applied to One of the Individual Cantilevers at the Self-Oscillating Frequency of the the System The synchronization phenomenon is not restricted to an interaction between oscillators; it is The to synchronization notarestricted to ansystem interaction between oscillators; it as is possible synchronize anphenomenon external forceiswith self-sustained through a weak interaction possible to synchronize an external force with a self-sustained system through a weak interaction as well [3]. The previous sub-section shows that, applying an external force to the cantilevers (at their well [3]. The previous showsthe that, applying an external force to modal the cantilevers (atoftheir own individual driver),sub-section they can change self-oscillation frequency to the frequency the own individual driver), they can change the self-oscillation frequency to the modal frequency of actuated cantilever; otherwise, no changes are measured. This is not proper synchronization withthe an actuatedforce. cantilever; otherwise, no changes are measured. This is not proper synchronization with an external external force. synchronization with an external force, we proceed in a similar way to that used in the To achieve To achieve synchronization with an external we proceed in a but, similar waycase, to that used in previous sub-section. We apply a stimulus to one offorce, the individual drivers in this at the same the previous sub-section. We apply a stimulus to one of the individual drivers but, in this case, at the same frequency as the self-oscillation frequency, i.e., at the frequency of the peak number 3 (see Figure 2). We present the results in the case of stimulating the cantilever number 2 (although similar results were obtained with the stimulus applied to cantilever number 4). Figure 9 depicts the frequency evolution over time, showing stabilization only when the stimulus is at the same

Sensors 2016, 16, 1690

9 of 15

frequency as the self-oscillation frequency, i.e., at the frequency of the peak number 3 (see Figure 2). Sensors 2016, 16, 1690 of 15 We present the1690 results in the case of stimulating the cantilever number 2 (although similar results 99were Sensors 2016, 16, of 15 obtained with the stimulus applied to cantilever number 4). Figure 9 depicts the frequency evolution frequency as the the self-oscillation self-oscillation (curves A and and B in in Figure Figure 9a). For instance, from Figure 9a, the the time time frequency as (curves A B 9a). instance, from 9a, over time, showing stabilization only when the stimulus is at theFor same frequency asFigure the self-oscillation frequency stability corresponding to case A is around 0.003 Hz/s; for case B it is 0.08 Hz/s and for frequency stability corresponding to case A is around 0.003 for case Bstability it is 0.08corresponding Hz/s and for (curves A and B in Figure 9a). For instance, from Figure 9a, theHz/s; time frequency −8 case C it it is 4.6 10−80.003 Hz/s.Hz/s; This fact fact is highlighted highlighted in Figure 9b, scaling the frequencies by the theThis starting case C ×× 10 Hz/s. This is Figure by starting to case Ais is4.6 around for case B it is 0.08 in Hz/s and9b, forscaling case C the it isfrequencies 4.6 × 10−8 Hz/s. fact point. point. is highlighted in Figure 9b, scaling the frequencies by the starting point.

(a) (a)

(b) (b)

Figure 9. 9. (a) Closed-loop Closed-loop frequency for for stimulus at at cantilever cantilever number number 22 for for different different frequencies: frequencies: A, A, Figure Figure 9. (a) (a) Closed-loopfrequency frequency forstimulus stimulus at cantilever number 2 for different frequencies: at 1.04 1.04 MHz; MHz; B, B, at at 1.06 MHz; MHz; C, C, at at 1.103552 1.103552 MHz; MHz; (b) Scaled Scaled frequencies by by the starting starting point for for the at A, at 1.04 MHz; B, 1.06 at 1.06 MHz; C, at 1.103552 MHz;(b) (b) Scaledfrequencies frequencies bythe the starting point point for the the same cases than in (a). same cases cases than than in in (a). (a). same

In Figure Figure 10 10 we we can can see see the the Allan Allan deviation deviation acting acting over over cantilever cantilever number number 22 with with different different In In Figure 10 we can seetothe Allan deviation acting over cantilever number 2 withWhen different frequencies corresponding the modal frequency of each of the five cantilevers. the frequencies corresponding to the modal frequency of each of the five cantilevers. When the frequencies corresponding to the modal frequency of each of the five cantilevers. When the frequency frequency of of the the stimulus stimulus takes takes the the same same value value as as the the self-oscillation self-oscillation frequency frequency or or the the modal modal frequency of the stimulus takes the same value as thethe self-oscillation frequency or the modal frequency of frequency of the corresponding cantilever, Allan deviation has the minimum value (curves A frequency of the corresponding cantilever, the Allan deviation has the minimum value (curves A the corresponding cantilever, the Allan deviation has the minimum value (curves A and B of and B B of of Figure Figure 10). 10). If If we we measure measure the the frequency frequency dispersion dispersion for for 11 ss of of averaging averaging time time for for the the and Figure 10). If we measure the frequency dispersion for 1Figure s of averaging time for the stimulation stimulation at the self-oscillation frequency (curve A in 10), we obtain ∆f = 0.013 Hz, and stimulation at the self-oscillation frequency (curve A in Figure 10), we obtain ∆f = 0.013 Hz, and at the self-oscillation frequency A in Figure 10), wemass obtain 0.013 using Equationthe (1) using Equation (1) (1) we we obtain (curve minimum detectable of∆f 2.6= ag. ag. AsHz, weand said previously, using Equation obtain aa minimum detectable mass of 2.6 As we said previously, the we obtain a minimum detectable mass of 2.6 ag. As we said previously, the minimum detectable mass minimum detectable detectable mass mass calculated calculated here here (using (using Equation Equation (1)) (1)) is is aa lower lower limit limit (see (see Section Section 4.2 4.2 for for minimum calculated here (using Equation (1)) is a lower limit (see Section 4.2 for more details). more details). more details).

Figure 10. Allan deviation, exciting cantilever number at different frequencies frequencies corresponding to the the number 22 at corresponding to Figure 10. Allan Allandeviation, deviation,exciting excitingcantilever cantilever number 2 different at different frequencies corresponding to different peaks of Figure 2a: A, peak number 3 (1.10355 MHz); B, peak number 2 (1.05494 MHz); C, the different peaks of Figure A, peak number 3 (1.10355 MHz); B, peak number 2 (1.05494 MHz); different peaks of Figure 2a: 2a: A, peak number 3 (1.10355 MHz); B, peak number 2 (1.05494 MHz); C, peak number 1 (1.06473 MHz); D, peak number 4 (1.0733 MHz); E, peak number 5 (1.09819 MHz). C, peak number 1 (1.06473 MHz); peak number 4 (1.0733 MHz); peak number 5 (1.09819 MHz). peak number 1 (1.06473 MHz); D, D, peak number 4 (1.0733 MHz); E, E, peak number 5 (1.09819 MHz).

Table 11 summarizes summarizes the the frequency frequency dispersion dispersion considering considering the the different different synchronization synchronization Table techniques. Note Note that that the the synchronization synchronization allows allows aa 23-fold 23-fold increase increase in in the the frequency frequency dispersion dispersion for for 11 techniques. s of averaging time. s of averaging time.

Sensors 2016, 16, 1690

10 of 15

Table 1 summarizes the frequency dispersion considering the different synchronization techniques. Note that the synchronization allows a 23-fold increase in the frequency dispersion for 1 s of averaging time. Table 1. Frequency dispersion for different averaging times. Measure Type

Frequency Dispersion at 1 s Averaging Time

Frequency Dispersion at 100 s Averaging Time

Without stimulation

0.3 Hz

6.8 Hz

Stimulation at cantilever number 2 at its modal frequency

0.016 Hz

0.004 Hz

Stimulation at cantilever number 4 at its modal frequency

0.017 Hz

0.0011 Hz

Stimulation at cantilever number 2 at the self-oscillation frequency

0.013 Hz

0.0013 Hz

4. Discussion 4.1. Considerations about Synchronization Using an External Force In this work, we present two ways of synchronizing the cantilevers: (a) exciting one of the cantilevers at its modal frequency and (b) exciting it at the self-oscillation frequency (the frequency in which the system oscillates without external stimulus on cantilevers). In both schemes we achieve the synchronization state using an external force. We discuss in this section the role of the external force as an agent for promoting synchronization due to the possibility that this external force can be synchronized with the system. Normally, one of the ways to study the synchronization between a self-sustained system and an external force is through the so-called Arnold tongue [3]. The Arnold tongue shows the region of frequencies in which the system is synchronized with the external force by plotting ∆F versus f, where ∆F is the difference between the frequency of the self-sustained system in the presence of the external force and the frequency of the external force, and f is the frequency of the external force. When the synchronization is achieved, a plateau appears, centered at the frequency of the self-sustained system without the presence of the external force. This plateau is wider as higher is the power of the external force. A 3D plot of ∆F versus f and versus power of the external force is the Arnold tongue (see, for example, [10]). Our second method of excitation (method (b)) can be represented using the Arnold Tongue, as can be seen in Figure 11. Figure 11a shows the synchronized zone, a shaded grey zone, which is a plateau that widens with the increasing of the excitation power (excitation amplitude in the y-axis of Figure 11). Figure 11b is a 2D representation of the Arnold Tongue that clarifies the effect of the excitation power over the synchronized zone. Red squares represent synchronized states and blue squares unsynchronized ones. Figure 12a shows a different representation of the Arnold tongue, adapted to our first method to achieve the synchronization (method (a)) at different amplitudes of the external force. To do this representation, we have done the measures using a frequency counter and applied the excitation (in Figure 12 the excitation is applied at cantilever number 4) at different frequencies. The z-axis represents the measure of the frequency counter. When we apply the external force with the same frequency as the modal frequency of the cantilever, the self-oscillation frequency changes from that corresponding with peak number 3 (at which frequency the system oscillates without external forcing; see Figure 2) to the modal frequency of the individual cantilever. In Figure 12a the synchronization zone appears as a plateau with the ratio between self-oscillation and excitation frequency equal to 1, and a different value for the out-of-synchronization zone (which corresponds to the ratio between the frequency of peak number 3 (self-oscillation frequency) and the frequency of the external force). The synchronization zone is centered at the modal frequency of cantilever number 4 and the range of excitation frequencies in which we have a synchronized state widens when the amplitude of excitation

Sensors 2016, 16, 1690 Sensors 2016, 16, 1690 Sensors 2016, 16, 1690

11 of 15 11 of 15 11 of 15

grows. Importantly, in the case of Figures 11a (see and Figure 12a, using of excitation amplitude there is a measures to obtain the time response signal 5a),350 wemV obtained a modulated signal at 350 measures to obtain the time response signal (see Figure 5a), we obtained a modulated signal at thin synchronized zone. When we performed our first measures to obtain the time response signal mV of excitation amplitude, which does not correspond to a synchronized state. This discrepancy350 is mV to of the excitation amplitude, does not correspond to aofperformed synchronized state. Thiswhich discrepancy is (see Figure 5a),that we obtained awhich modulated signal at 350 were mV excitationby amplitude, does due fact the time response signal measures applying 18 V DC at not the due to the fact the time signal measures performed by applying 18 Vthan DCsignal at the the correspond to athat synchronized state. discrepancy iswere due to the fact24that the time response closed-loop configuration, butresponse in the This last case the applied voltage was V—more energy in closed-loop configuration, but in the last case the applied voltage was 24 V—more energy than in the measures were performed by applying 18 V DC at the closed-loop configuration, but in the last case first case, which caused the early appearance of the synchronized state. first case, which caused the early appearance of the synchronized state. the applied was 24 V—more energy in the first case, which caused theintuitive early appearance Figure voltage 12b represents the widening ofthan the synchronization zone in a more manner, 12bsynchronized represents widening of the synchronization zone in a more intuitive manner, of theFigure synchronized state. the capturing the zone with red squares and the non-synchronized zone with blue capturing the synchronized zone with red squares and the non-synchronized zone with blue Figure 12b represents the widening of the synchronization zone in a more intuitive manner, squares. squares. the synchronized zone with red squares and the non-synchronized zone with blue squares. capturing

(a) (a)

(b) (b)

Figure 11. (a) The Arnold tongue, corresponding to excitation at the same frequency as the Figure 11. 11. (a) The tongue, to excitation the same frequency as the the self-sustained (method (b)). The shadedcorresponding gray zone is the which frequency widens when Figure (a) The Arnold Arnold tongue, corresponding tosynchronized excitation at atzone, the same as the self-sustained (method (b)). The shaded gray zone is the synchronized zone, which widens when the amplitude of excitation grows; (b) A 2D representation of the Arnold tongue, whichwhen the red self-sustained (method (b)). The shaded gray zone is the synchronized zone, whichin widens the amplitude excitation grows; A representation 2D representation ofArnold the Arnold tongue, in and which the red squares represent the grows; minimum between self-oscillation frequency excitation amplitude ofofexcitation (b)(b) A difference 2D of the tongue, in which the red squares squares represent the difference minimumbetween difference between self-oscillation frequencyfrequency. and excitation frequency. represent the minimum self-oscillation frequency and excitation frequency.

1.03 1.03

1.02 1.01 1.01 1.00

700 600 700 500600 400500

1.00 1.070

1.072 Exc 1.070 itatio 1.074 n Fr 1.072 equ Exc enc itatio 1.074 y( n

Ex Exc ci ta itati tio on n Am Am pl plit itu ud de e (m (mV V) )

/ fexc f xc f /elfe fselfs

1.02

1.076

MHz Freq ) 1.076 uen cy ( MHz )

(a) (a)

300 400 300

(b) (b)

Figure 12. (a) self-oscillation and and excitation excitation frequencies frequencies (fself (fself/fexc) Figure 12. (a) Ratio Ratio between between self-oscillation /fexc) for for different different Figure 12. of (a) the Ratio between self-oscillation and excitation frequencies (fself/fexc) isforapplied different amplitudes excitation and 24 V DC of applied voltage. Here, the excitation amplitudes of the excitation and 24 V DC of applied voltage. Here, the excitation is applied at at amplitudes of the 4.excitation and 24 Vzone DC appears of applied voltage. Here, the excitation is applied at cantilever number The synchronized as a plateau for which the ratio fself/fexc cantilever number 4. The synchronized zone appears as a plateau for which the ratio fself /fexc takes takes cantilever number 4. The synchronized zone appears as a plateau for which the ratio fself/fexc takes the the unit unit value value (shaded (shaded gray gray zone); zone); (b) (b) A A 2D 2D representation representation of of the the previous previous figure, figure, in in which which the the red red the unitrepresent value (shaded gray zone); (b) A 2D representation of thethe previous figure, incan which the red squares the minimum ratio of fself/fexc and the blue ones maximum. We see that squares represent the minimum ratio of fself /fexc and the blue ones the maximum. We can see that the the squares represent the minimum ratio of and the blue the maximum. can see that the rage ofexcitation excitationfrequencies frequencies thefself/fexc synchronized zoneones widens the We amplitude of the rage of intointo the synchronized zone widens with thewith amplitude of the excitation. rage of excitation frequencies into the synchronized zone widens with the amplitude of the excitation. excitation.

In the case of applying the external force with the same frequency at which the self-sustained In the case of applying the external force with the same frequency at which the self-sustained system oscillates (the frequency of the self-oscillation in a closed-loop we conclude In oscillates the case of applying the of external force with the frequencyconfiguration), at which the self-sustained system (the frequency the self-oscillation insame a closed-loop configuration), we conclude system (the frequency of the in a closed-loop weare conclude that the oscillates external force synchronizes withself-oscillation the system following the usualconfiguration), way [10] i.e., we on the that the external force synchronizes with the system following the usual way [10] i.e., we are on the

Sensors 2016, 16, 1690

12 of 15

that the external force synchronizes with the system following the usual way [10] i.e., we are on the Arnold tongue plateau. This synchronization drives our system to a more stable oscillation frequency system than without the applied external force. We have to take into account that, when we use the closed-loop configuration, the frequency of oscillation of the system is unique, corresponding to the frequency of peak number 3 (see Figure 2); we do not have multivalued frequency as is measured in thermomechanical noise. It is important to emphasize that the external force is synchronized with the whole system. When we apply an external force at the individual modal frequency of one of the five cantilevers, in our opinion, the way to achieve synchronization is not the same as in the previous case. In this configuration, the self-sustained system oscillates at a frequency (1.10355 MHz) that is very different to the modal frequency of one of the actuated cantilevers (for example, 1.07348 MHz for cantilever number 4). If we excite the system with an external force with the modal frequency of the corresponding actuated cantilever, we are exciting our system with a frequency far from the self-oscillation one, that is, we are out of the Arnold tongue plateau (showed in Figure 11a), which corresponds to synchronization to an external force and, consequently, a synchronization state between external force and self-sustained system should not be achieved. However, as we see, the whole system changes its self-oscillation frequency from 1.103 MHz to, for example, 1.073 MHz (if we excite cantilever number 4) and we obtain a synchronized system with an Arnold tongue (like the one depicted in Figure 12a), but shifted to the modal frequency of the excited cantilever. As we can see in Figure 2, the fabrication process produces a system with five cantilevers with enough differences between their modal frequencies to disallow the mutual synchronization. In our opinion, the role of the external force using the same frequency as the modal of the individual cantilever is helping the mutual synchronization between cantilevers. When the external force is applied at the modal frequency of the actuated cantilever, the whole system oscillates at this frequency (see Figure 6). This fact drives us to discard the direct influence of the force on the frequency of oscillation but to assert an indirect influence only when its frequency and the modal one of the cantilevers matches. 4.2. Considerations about Mass Sensor Performance Even though this synchronizable system could work with different purposes, we want to analyze it for mass sensing applications. We have introduced two ways to achieve synchronization with the purpose of achieving higher frequency stability. However, there are some considerations to be taken into account, directly related to the fact that, in synchronous operation, the system is locked with an external force. The Arnold tongue shows as a plateau the region in which the system frequency and the external force frequency are locked. It is important to state that, in order to use this synchronized system as a mass sensor measuring frequency change, the added mass should be able to shift the frequency by more than half of the width of the Arnold tongue plateau. Consequently, working with the minimum achievable power of the external force compatible with synchronization, i.e., the narrower zone of the Arnold tongue, is needed for mass sensing. In the case of an external force applied on one of the individual cantilevers, at the same frequency as the modal frequency of the corresponding cantilever, the spatial identification of the deposited mass over the five cantilevered array is achieved as is discussed below. Imagine we have a system vibrating synchronously at the same frequency as the modal of one of the cantilevers, forced by an external force (the case described in Section 3.1). When we deposit mass over this cantilever, its modal frequency changes in direct relation to the amount of mass and then a mismatch between the modal frequency of the cantilever and the frequency of the external force appears. Consequently (if the intensity of the external force is not too high), the synchronization disappears (the system moves out of the Arnold tongue). Based on this fact, we will be able to identify which of the cantilevers has the added mass exciting sequentially each cantilever at its corresponding modal frequency, while deciding if the system works synchronously or not. The cantilever for which this procedure does not lead the system to work synchronously will be the cantilever where the mass

Sensors 2016, 16, 1690

13 of 15

has been deposited. This is an advantage due to the fact that we can perform a spatial detection of mass deposition. For instance, this method can be useful for profiling a flux of particles. Moreover, this method will allow the chemical identification of the deposited mass if we use a different and specific chemical functionalization of each cantilever in the array (for example, viruses caught by antigens sited at the corresponding cantilever, which will be spatially determined by losing the synchronization state). The disadvantage of this method of achieving synchronization is that, directly, we cannot measure the mass deposited because we cannot measure the shift of frequency (i.e., Equation (1) is not applicable in this case). To know the mass we have to perform another step. Centered at the identified cantilever (the cantilever with the added mass), we have to change the frequency of the external force until the system returns to a synchronous state. The difference between this new frequency and the previous one (the modal frequency of the cantilever without added mass) will determine the mass deposited at the cantilever. It is important to state that if we want to achieve the minimum detectable mass, we have to use the minimum external force to have the minimum range of synchronous state (the narrower zone of the Arnold tongue). This procedure requires a previous calibration of the system and two steps of excitation detection. 4.3. Towards a Thermomechanical Noise Limit? Recently there has been an exhaustive review of the frequency stability of micro and nanomechanical resonators [2]. This review states that the frequency stabilities of the NEMS resonators studied are far from the thermomechanical limit and none of them attain this limit. The authors discard the idea that this discrepancy between measurements and thermomechanical limit is due to the measurement system; rather, it originated in the mechanical domain of the device. The authors also disagree with the idea that the difference between the measured frequency fluctuations and the thermomechanical limit is due to temperature variations and another known mechanisms and conclude that there is a need for studying new microscopic mechanisms, which might be the origin of these discrepancies. According to this paper, the measured frequency fluctuation is on average two orders of magnitude (100-fold) greater than the thermomechanical frequency fluctuation limit. We can evaluate in the same terms as [2] our CMOS-MEMS cantilevered array’s ability to compare stability with a synchronized operation mode and without. For our system the thermomechanical Allan deviation limit is computed to be 4.6 × 10−9 with 1 s averaging time (see the supplementary information in [2] for how to compute it). From Figure 10, the Allan deviation is 3.5 × 10−7 (@ τ = 1 s) without synchronized operation, which represents a factor 80x greater than the thermomechanical frequency stability in accordance with predictions and results in [2]. On the other hand, the Allan deviation becomes 1.3 × 10−8 (@ τ = 1 s) in a synchronized mode of operation, which represents an increase of only 2.6x in comparison with the equivalent thermomechanical limit, surpassing in this way the expected frequency stability for the majority of NEMS resonators. This opens the way to further push the limits for hig- performance mass sensors. Additionally, in our opinion, synchronized systems are revealed as a good strategy to attain the thermomechanical limit, and this fact must be taken into account to face the research concerning the new microscopic mechanism proposed by the authors in [2]. 5. Conclusions We present here a multicantilevered system with enhanced frequency stabilization of the self-oscillation frequency through synchronization. We have concluded that the frequency stability improves by around two orders of magnitude using synchronization phenomena. We present two ways to synchronize the cantilevers, both using an external force: one of them is to use this external force as a way to overcome the difficulties related to the dispersion of dimensions of the cantilevers due to the fabrication process, acting on an individual cantilever at its modal frequency and driving the whole system to oscillate at this frequency; the other is synchronizing the external force with the

Sensors 2016, 16, 1690

14 of 15

self-sustained system. For both synchronization methods we have presented the advantages and disadvantages using this system as a mass sensor. Acknowledgments: This work was supported by the Spanish Ministry of Economy and Competitiveness (MINECO), projects NemsInCMOS (TEC2012-32677) and CMOS-MENUTS (TEC2015-66337-R). Author Contributions: F.T. designed the system and measured and analyzed the results presented in this paper. A.U. contributed to design the system and perform the measurements. N.B. contributed to designing the system and analyzing the results. F.T. and N.B. have written the manuscript. M.R. and G.S. have contributed to sample preparation and characterization. Conflicts of Interest: The authors declare no conflict of interest.

Abbreviations The following abbreviations are used in this manuscript: CMOS M/NEMS TIA

Complementary Metal Oxide Semiconductor Micro/Nano ElectroMechanical Systems TransImpedance Amplifier

References 1. 2.

3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17.

Chaste, J.; Eichler, A.; Moser, J.; Ceballos, G.; Rurali, R.; Bachtold, A. A nanomechanical mass sensor with yoctogram resolution. Nat. Nanotechnol. 2012, 7, 301–304. [CrossRef] [PubMed] Sansa, M.; Sage, E.; Bullard, E.C.; Gély, M.; Alava, T.; Colinet, E.; Naik, A.K.; Villanueva, L.G.; Duraffourg, L.; Roukes, M.L.; et al. Frequency fluctuations in silicon nanoresonators. Nat. Nanotechnol. 2016, 11, 552–558. [CrossRef] [PubMed] Pikovsky, A.; Rosenblum, M.; Kurths, J. Synchronization, a Universal Concept in Nonlinear Sciences; Cambridge University Press: Cambridge, UK, 2001. Huygens, C. Instructions concerning the use of pendulum-watches for finding the longitude at sea. Philos. Trans. R. Soc. Lond. 1669, 4, 937–976. Lifshitz, R.; Kenig, E.; Cross, M.C. Collective Dynamics in Arrays of Coupled Nonlinear Resonators. In Fluctuating Nonlinear Oscillators; Oxford University Press: Oxford, UK, 2012; Chapter 11. Stankovski, T.; McKlintock, P.; Stefanovska, A. Dynamical interference: Where phase synchronization and generalized synchronization meet. Phys. Rev. E 2014, 89, 062909. [CrossRef] [PubMed] Buks, E.; Roukes, M.L. Electrically tunable collective response in a coupled micromechanical array. J. Microelectromech. Syst. 2002, 11, 802–807. [CrossRef] Sato, M.; Hubbard, B.E.; Sievers, A.J. Observation of locked intrinsic localized vibrational modes in a micromechanical oscillator array. Phys. Rev. Lett. 2003, 90, 044102. [CrossRef] [PubMed] Baghelani, M.; Ebrahimi, A.; Ghavifekr, H.B. Design of RF MEMS based oscillatory neural network for ultra high speed associative memories. Neural Process. Lett. 2014, 40, 93–102. [CrossRef] Shim, S.; Imboden, M.; Mohanty, P. Synchronized oscillation in coupled nanomechanical oscillators. Science 2009, 316, 95–99. [CrossRef] [PubMed] Feng, J.; Ye, X.; Esashi, M.; Ono, T. Mechanically coupled synchronized resonators for resonant sensing applications. J. Micromech. Microeng. 2010, 20, 115001. [CrossRef] Cross, M.C.; Zumdieck, A.; Lifshitz, R.; Rogers, J.L. Synchronization by nonlinear frequency pulling. Phys. Rev. Lett. 2004, 93, 224101. [CrossRef] [PubMed] Agrawal, D.K.; Woodhouse, J.; Seshia, A. Observation of locked phase dynamics and enhanced frequency stability in synchronized micromechanical oscillators. Phys. Rev. Lett. 2013, 111, 084101. [CrossRef] [PubMed] Agrawal, D.K.; Woodhouse, J.; Seshia, A. Synchronization in a coupled architecture of microelectromechanical oscillators. J. Appl. Phys. 2014, 115, 164904. [CrossRef] Thiruvenkatanathan, P.; Yan, J.; Woodhouse, J.; Seshia, A. Enhancing Parametric Sensitivity in Electrically Coupled MEMS Resonators. J. Micromech. Syst. 2009, 18, 1077–1086. [CrossRef] Zhao, C.; Wood, G.S.; Xie, J.; Chang, H.; Pu, S.H.; Kraft, M. A three degree-of-freedom weakly coupled resonator sensor with enhanced stiffness sensitivity. J. Microelectromech. Syst. 2016, 25, 38–51. [CrossRef] Almog, R.; Zaitsev, S.; Shtempluck, O.; Buks, E. Noise squeezing in a nanomechanical Duffing resonator. Phys. Rev. Lett. 2007, 98, 078103. [CrossRef] [PubMed]

Sensors 2016, 16, 1690

18.

19. 20. 21.

15 of 15

Villanueva, L.G.; Kenig, E.; Karabalin, R.B.; Matheny, M.H.; Lifshitz, R.; Cross, M.C.; Roukes, M.L. Surpassing fundamental limits of oscillators using nonlinear resonators. Phys. Rev. Lett. 2013, 110, 177208. [CrossRef] [PubMed] Kacem, N.; Baguet, S.; Duraffourg, L.; Jourdan, G.; Dufour, R.; Hentz, S. Overcoming limitations of nanomechanical resonatorswith simultaneous resonances. Appl. Phys. Lett. 2015, 107, 073105. [CrossRef] AMS. Available online: www.ams.com (accessed on 6 May 2016). Verd, J.; Uranga, A.; Abadal, G.; Teva, J.; Torres, F.; López, J.L.; Pérez-Murano, F.; Esteve, J.; Barniol, N. Monolithic CMOS MEMS oscillator circuit for sensing in the attogram range. IEEE Electron Device Lett. 2008, 29, 146–148. [CrossRef] © 2016 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC-BY) license (http://creativecommons.org/licenses/by/4.0/).