Enzyme-Linked Immunosorbent Assay for ... - Semantic Scholar

4 downloads 0 Views 75KB Size Report
The use of antibiotics in animal feed helps. (ELISA) kits that are commonly used for tylosin or tetracycline resi- increase the animal's ability to absorb feed and ...
Enzyme-Linked Immunosorbent Assay for Ultratrace Determination of Antibiotics in Aqueous Samples Kuldip Kumar,* Anita Thompson, Ashok K. Singh, Yogesh Chander, and Satish C. Gupta ABSTRACT

K. Kumar, Y. Chander, and S.C. Gupta, Department of Soil, Water, & Climate, and A.K. Singh, Department of Veterinary Diagnostic Medicine, University of Minnesota, St. Paul, MN 55108. A. Thompson, Department of Biological Systems Engineering, University of Wisconsin, Madison, WI 53706. Received 20 Mar. 2003. *Corresponding author ([email protected]).

ronmental Media Services (2000), more than 40% of the antibiotics produced in the U.S. are used as feed supplements. The use of antibiotics in animal feed helps increase the animal’s ability to absorb feed and reach market weight on time. It may also counteract the effects of crowded living conditions and poor hygiene in intensive animal agriculture systems (Environmental Media Services, 2000). Antibiotics commonly used as feed additive for animals include chlortetracycline, bacitracin, bambermycins, erythromycin, lincomycin, monensin, oleandomycin, oxytetracycline, penicillin, tylosin, and virginiamycin (Church and Pond, 1982). The antibiotic dose varies from 1 to 100 g Mg⫺1 of feed depending upon the type and size of the animal and the type of antibiotic. Most of the antibiotics added to animal feed are excreted in urine or feces. In some cases, as much as 80% of the antibiotic administered orally may pass through the animal unchanged (Levy, 1992). Once excreted in urine and feces, antibiotics may enter surface and/or ground waters through nonpointsource pollution from manure-applied lands. Land application of manure is a common practice in many parts of the USA. In the northern tier of the country, manure is applied even during winter over snow. Manure is applied to land because of its value in supplying nutrients to crops as well as a means of disposing unwanted waste. Recently, the United States Geologic Survey (Koplin et al., 2002) reported the presence of several antibiotics in 139 streams across 30 states in the United States. However, the contribution of agricultural runoff as compared with wastewater (from sewage treatment plants) in terms of antibiotics is unclear. The conventional trace analysis methods for tylosin and tetracycline depend on high performance liquid chromatography (HPLC) or liquid chromatography– mass spectrometry (LC–MS) (Cooper et al., 1998; Loke et al., 2000; Sacher et al., 2001). However, these methods are expensive and time consuming. Recently, enzymelinked immunosorbent assays (ELISA) have been shown to be useful for analyzing antibiotic residues in milk, meat, fish, eggs, and honey (Lee et al., 2001; De Wasch et al., 2001; Draisci et al., 2001; Mascher et al., 2001). This technique has only recently been introduced into environmental chemistry research for the determination of herbicides in surface and ground water samples (Thurman et al., 1990; Aga and Thurman, 1993). Meyer et al. (2000) developed a radioimmunoassay procedure for quantification of tetracycline antibiotics in wastewaters; however, the concentrations these authors measured using this procedure were substantially less than

Published in J. Environ. Qual. 33:250–256 (2004).  ASA, CSSA, SSSA 677 S. Segoe Rd., Madison, WI 53711 USA

Abbreviations: ELISA, enzyme-linked immunosorbent assay; LC– MS, liquid chromatography–mass spectrometry; OD, optical density.

Two commercially available enzyme-linked immunosorbent assay (ELISA) kits that are commonly used for tylosin or tetracycline residues in meat and milk were adapted for ultratrace analysis of these antibiotics in surface and ground waters. These two antibiotics are commonly fed to swine, turkeys, and cattle at subtherapeutic doses for growth promotion purposes. Both ELISA techniques were found to be highly sensitive and selective for the respective antibiotics with detection limits of 0.10 and 0.05 ␮g L⫺1 for tylosin and tetracycline, respectively. The recovery of both tylosin and tetracycline from spiked samples of lake waters, runoff samples, soil saturation extracts, and nanopure water was close to 100%. Tetracycline ELISA was highly specific for tetracycline and chlortetracycline but not for other forms of tetracycline (oxytetracycline, demeclocycline, and doxycycline). Analysis of a few liquid swine manure samples by liquid chromatography–mass spectrometry (LC–MS) showed lower concentrations for chlortetracycline as compared with concentrations obtained using ELISA. However, the concentrations of tylosin from ELISA were comparable with that of LC–MS. The lower concentrations of chlortetracycline obtained by LC–MS in manure samples indicate the presence of other similar or transformed compounds that were detected by ELISA but not determined by LC–MS. These results indicate that both ELISA kits can be useful tools for low-cost screening of tylosin, tetracycline, and chlortetracycline in environmental waters. Furthermore, both ELISA procedures are rapid, portable, and easily adaptable for testing of multiple samples simultaneously.

S

ince their discovery, antibiotics have been instrumental in treating infectious diseases that were previously known to kill humans and animals. However, it has now become clear that widespread use of antibiotics is not without problems (Halling-Sørensen et al., 1998; Jørgensen and Halling-Sørensen, 2000). The major concern is the development of antibiotic-resistant microorganisms, which are difficult to treat with existing antibiotics (Ford, 1994; Herron et al., 1997). Increasingly more microorganisms are becoming resistant to multiple antibiotics (Goldburg, 1999). According to one estimate, more than 22.7 million kg of antibiotics are currently being produced each year in the USA (Environmental Media Services, 2000). Although most of these antibiotics are used for treatment of infections in humans and animals, a significant portion is used as a supplement in animal feed to promote growth of food-producing animals. According to Envi-

250

KUMAR ET AL.: ELISA ASSAY FOR ANTIBIOTIC DETERMINATION

those measured using LC–MS. The advantages of ELISA compared with HPLC and LC–MS have been recognized by the USEPA (1992) and guidelines for evaluating ELISA kits are being developed. This study examined the use of ELISA method for measuring the concentration of tylosin- and tetracycline-class antibiotics in water samples. These two antibiotics are routinely fed to animals and are often present in manure (Kumar et al., 2002). The specific objective of this study was to evaluate two commercially available ELISA kits (used to test antibiotic residues in milk, meat, or honey samples) for measuring tylosin and tetracycline concentrations in runoff and percolating waters from manure-applied fields and surface water bodies. We used soil saturation extracts to simulate percolating waters. MATERIALS AND METHODS Reagents Pure antibiotic salts of tylosin tartrate, tetracycline hydrochloride, chlortetracycline hydrochloride, oxytetracycline hydrochloride, doxycycline hydrochloride, and demecocycline hydrochloride were purchased from Sigma Chemicals (St Louis, MO). A tylosin one-step ELISA kit was obtained from International Diagnostic Systems Corporation (St. Joseph, MI) and Ridascreen tetracycline kits were obtained from R-Biopharm AG (Darmstadt, Germany).

Enzyme-Linked Immunosorbent Assay Basis Tylosin The tylosin test is based on solid phase immunoassay technology. Tylosin-containing standards or water samples are added to microtiter wells coated with high affinity capture antibody to tylosin. The tylosin enzyme conjugate competes with tylosin in the sample for binding sites on the capture antibody. After a wash step, a substrate is added that reacts to any bound enzyme, creating a blue color. Tylosin in samples blocks the binding of enzyme conjugate to the capture antibody, resulting in little or no color development depending on the amount of tylosin in the sample. Results are quantified by measuring optical density (OD) (at 450 nm) of both standards and samples after stopping the reaction with a stop solution in a microplate reader. The OD is inversely proportional to tylosin concentration in the sample. Tetracycline The basis of the tetracycline ELISA test is similar to that of tylosin except that it is an antigen–antibody reaction. The microtiter wells coated with tetracycline–protein conjugate are mixed with tetracycline containing standards or water samples. Free tetracycline and immobilized tetracycline compete for the tetracycline antibody binding sites (competitive enzyme immunoassay). Any unbound antibody is then removed in a washing step and enzyme-labeled secondary antibody (which was directed against the anti-tetracycline antibody) is added. After removing unbound enzyme-labeled antibodies with a wash step, enzyme substrate (urea peroxide) and chromogen (tetramethyl-benzidine) are added to the wells and incubated. Bound enzyme conjugate converts the colorless chromogen into a blue product. The addition of the stop reagent leads to a color change from blue to yellow. Results are quantified by measuring OD of both standards and samples at 450 nm in a

251

microplate reader. The OD is inversely proportional to tetracycline concentration in the sample.

Analytical Procedures Tylosin Twenty microliters each of six standards ranging from 0 to 20 ␮g of tylosin L⫺1 and a sample diluent as a negative control were added in microtiter wells of the ELISA plate. The samples were diluted with sample diluent buffer (supplied with the kit) in the ratio of 1:1 before adding them to the microtiter wells. One hundred microliters of reconstituted tylosin enzyme was added to each well and incubated on a plate shaker for 10 min at room temperature. The solution in the plate wells was dumped out by inverting and blotting on paper towels. Using a multichannel pipette, 400 ␮L of wash solution was then added and the wells were inverted immediately and tamped on paper towels three or four times to remove excess moisture. The wash step was repeated three times. If bubbles were present, a clean pipette tip was used to break them. The wells were not allowed to dry and 150 ␮L of one-piece substrate was added to each well. The substrate was allowed to react at room temperature by placing on a plate shaker for 10 min. A medium blue color developed in the wells with the negative control and samples with no tylosin. The reaction was stopped by adding 150 ␮L of stop solution (supplied with kit) to each well. Using a microplate reader and a 450-nm filter, the OD of each well was read. Optical density is inversely related to tylosin concentration. Using a standard curve and appropriate sample dilution factor, the concentration of tylosin in water samples was determined. Analysis of each water sample, standards, and control was replicated three times. Tetracycline Fifty microliters each of prepared water samples (1:1 dilution with same buffer used to make standards) and standards were added in microtiter wells of the ELISA plate. Fifty microliters of anti-tetracycline antibody solution was then added to each well. The plates were incubated on a plate shaker for 1 h at room temperature. The solution in the wells was discarded and the microplate was tapped three or four times on blotting paper to ensure complete removal of solution from wells. The wells were filled with 250 ␮L of washing buffer. The liquid was again poured out and the wash step was repeated three times. One hundred microliters of enzyme conjugate was then added to each well, mixed gently on a plate shaker, and incubated at room temperature for 15 min. The liquid in the wells was discarded and the wash step was repeated three times. Next, 50 ␮L of substrate and 50 ␮L of chromogen were added to each well. Samples were mixed on a plate shaker and then incubated again for 15 min. Finally, 100 ␮L of stop solution was added and the samples were mixed on a plate shaker. The OD was measured at 450 nm. Analysis of each sample was replicated three times.

Verification of Enzyme-Linked Immunosorbent Assay Selectivity Selectivity is the ability of a test to determine that negative samples are truly negative. The selectivity test was performed on saturation extracts of four different soils, surface water from two different lakes, runoff water from two agricultural fields that had no history of manure application, and a sample of nanopure water. The soils used in this study were: Seelyeville muck (euic, frigid Typic Haplosaprists) from Randall,

252

J. ENVIRON. QUAL., VOL. 33, JANUARY–FEBRUARY 2004

Table 1. Optical density (OD) and the corresponding inhibition values for negative control (buffer) and different background matrices using the tylosin and tetracycline enzyme-linked immunosorbent assay (ELISA) kits. Tylosin ELISA Negative control (buffer) OD Mean Standard deviation CV

1.70 0.19 0.11

Different background matrices†

Tetracycline ELISA Standard (0.25 ␮g L⫺1)

⌬OD‡

Inhibition value§

⌬OD

Inhibition value

Negative control (buffer) OD

0.09 0.01 0.10

7.00 0.50 0.07

0.45 0.07 0.15

26.00 2.20 0.08

1.90 0.20 0.10

Different background matrices

Standard (0.15 ␮g L⫺1)

⌬OD

Inhibition value

⌬OD

Inhibition value

0.10 0.01 0.09

5.00 0.20 0.04

0.40 0.08 0.06

23.00 0.92 0.04

† Different background matrices consisted of soil saturation extract, runoff water, and water samples from lakes. ‡ ⌬OD ⫽ maximum OD in negative control minus the OD of zero concentration. § Inhibition value ⫽ [(OD negative control ⫺ OD sample) ⫻ 100]/OD negative control.

MN; Hubbard loamy sand (sandy, mixed, frigid Entic Hapludolls) from Becker, MN; Webster clay loam (fine-loamy, mixed, superactive, mesic Typic Endoaquolls) from Lamberton, MN; and Rozetta silt loam (fine-silty, mixed, superactive, mesic Typic Hapludalfs) from Lancaster, WI. All samples except nanopure water were filtered through 0.7-mm glassfiber filters (GF/F; Whatman, Maidstone, UK) before any testing. In addition to the above samples, a set of samples to be tested for tylosin were spiked with 5 ␮g L⫺1 of tetracycline and another set to be tested for tetracycline was spiked with 5 ␮g L⫺1 of tylosin. These samples were spiked to test the cross-reactivity of these antibiotics. All selectivity tests were performed in triplicate.

Enzyme-Linked Immunosorbent Assay Sensitivity The same matrices used to test ELISA selectivity were also spiked with three or four different concentrations (above the detection limits) to test the sensitivity of the ELISA kits. These tests were also performed in triplicate.

Liquid Manure Sample Preparation for Enzyme-Linked Immunosorbent Assay and Liquid Chromatography–Mass Spectrometry Four swine manure samples were obtained from the hog farmers in Minnesota. Twenty milliliters of liquid manure was prepared for extraction by adding 50 ␮L of 40% H2SO4, 30 mL nanopure water, and a 1-g scoop of disodium ethylenediaminetetracetate (Na2EDTA). The samples were placed on an orbital shaker for 4 h at 100 rpm. After shaking the samples were filtered using 0.7-␮m glass-fiber filters. A small aliquot of the filtered sample was further diluted (1:100 to 1:250) with dilution buffer supplied with ELISA kits. The diluted samples were analyzed using the ELISA procedure described earlier. For LC–MS analysis, the rest of the filtered sample was processed using solid-phase extraction (SPE) with 60-mg hydrophilic–lipophilic (HLB) cartridges from Waters (Milford, MA) using the procedure described by Lindsey et al. (2001). The eluted samples from HLB cartridges were then analyzed by LC–MS.

Liquid Chromatography–Mass Spectrometry Analysis The procedure for the confirmation of the chlortetracycline and tylosin using LC–MS was adapted from a method described by Zhu et al. (2001). Isocratic separation was achieved using a 150- ⫻ 4.6-mm C18 column. Ion source and trap conditions were adjusted and programmed to achieve the most stable and intense product ions that provided maximum sensitivity for each compound. The conditions were the following: ESI source voltage ⫽ 4.03 kV, vacuum ⫽ 1.02 ⫻ 10⫺5 Torr, lens voltage ⫽ 25.73 V, and multiplier voltage ⫽ 973.5 V. The LC–MS was programmed to monitor ions from m/z 100 to 1100.

RESULTS Enzyme-Linked Immunosorbent Assay Sensitivity

Fig. 1. Relationship between percent inhibition versus (a ) log tylosin concentration and (b ) log tetracycline concentration.

There was substantial variability (data not reported) in OD (absorbance units) between the negative control (dilution buffer) supplied with the kits and different background matrices (surface water, runoff water, nanopure water, and the soil saturation extracts). This variability reduced to 5 to 7% when dilution buffer was

253

KUMAR ET AL.: ELISA ASSAY FOR ANTIBIOTIC DETERMINATION

Table 2. Selectivity of enzyme-linked immunosorbent assay (ELISA) tests for tylosin and tetracycline for different background aqueous matrices (antibiotic negative samples tested negative using ELISA).

Sample matrix

Number of samples

Soil saturation extracts Surface waters Runoff waters Spiked tylosin in nanopure water Spiked tetracycline in nanopure water Total

12 8 10 5 5 35

Tetracycline ELISA negative samples

Tylosin ELISA negative samples Number

Percentage

Number

Percentage

11 8 10 NT† 5 34

92 100 100 NT 100 99

11 8 9 5 NT 33

100 100 90 100 NT 97

† Not tested.

added to various background matrices in 1:1 volume and the OD was expressed as inhibition value (normalized difference in OD of the sample with respect to the negative control): Inhibition value ⫽ [(OD of negative control ⫺ OD of sample)/OD of sample] ⫻ 100 The data in Table 1 show that the coefficient of variation in OD values for the tylosin and tetracycline ELISA tests varied from 10 to 15 and 6 to 9%, respectively. However, when inhibition was calculated this variation reduced to 7 to 8 and 4% for tylosin and tetracycline, respectively. The 35 blank samples from 6 different matrices had an inhibition value of 7% with a standard deviation of 0.5 for tylosin ELISA and 5% inhibition value with a standard deviation of 0.2 for tetracycline ELISA (Table 1). Taking into consideration the recommendations of the Subcommittee on Analysis (American Chemical Society Committee on Environmental Improvement and Subcommittee on Environmental Analytical Chemistry, 1980), the limits of detection were set at three standard deviations above the blank. In this case the detection limits corresponded to 9% inhibition for tylosin and 6% inhibition for tetracycline. These numbers correspond to detection limits of 0.10 ␮g L⫺1 for tylosin and 0.05 ␮g L⫺1 for tetracycline in water

samples. Because of the reduction in variability on dilution buffer addition, all water samples were mixed with dilution buffer (1:1) before analysis. The standard curves of inhibition value against log of antibiotic concentration were linear from 0.25 to 5.00 ␮g L⫺1 for tylosin ELISA and from 0.15 to 4.05 ␮g L⫺1 for tetracycline ELISA (Fig. 1).

Enzyme-Linked Immunosorbent Assay Selectivity Both tetracycline and tylosin ELISA procedures were highly selective for their respective antibiotics (Table 2). The mean (n ⫽ 35) selectivity of tylosin ELISA was 99% and that of tetracycline ELISA was 97%. Also, there was no cross-reactivity of tylosin and tetracycline as tylosin-spiked sample showed negative results with tetracycline ELISA and vice versa (Table 2).

Recovery Recovery of tylosin from spiked samples of various background matrices was generally between 95 and 110% with a mean recovery of 99% (Table 3). There were two samples (out of 21 samples) where tylosin recovery was only 88 and 92%. Similarly, the recoveries of tetracycline varied between 95 and 122% with a mean

Table 3. Concentration of tylosin in different aqueous samples obtained with tylosin enzyme-linked immunosorbent assay (ELISA) kit. Matrix Minnehaha Creek (Date 1)

Minnehaha Creek (Date 2) Silver Lake Runoff water Soil saturation extract Hubbard loamy sand Webster clay loam Seeleville muck Rozetta silt loam

Spiked concentration of tylosin

Dilution factor†

Mean tylosin concentration using ELISA

Mean percent recovery

␮g L⫺1 2.0 4.0 15.0 40.0 7.5 15.0 40.0 1.0 5.0 10.0 2.5 5.0 10.0

2 2 5 20 3 5 20 2 2 5 2 2 5

␮g L⫺1 2.1 4.3 14.6 37.0 7.5 15.0 38.0 1.1 4.4 9.8 2.6 4.8 10.1

% 105 107 97 92 100 100 95 110 88 98 104 96 101

5.0 10.0 5.0 10.0 5.0 10.0 5.0 10.0

2 5 2 5 2 5 2 5

5.2 10.0 5.1 10.0 4.9 9.9 5.0 9.9

104 100 102 100 98 99 100 99

† The samples were diluted using dilution buffer supplied with the kit to bring the concentration within linear part of standard curve (0.25–5 ␮g L⫺1).

254

J. ENVIRON. QUAL., VOL. 33, JANUARY–FEBRUARY 2004

Table 4. Concentration of tetracycline in different aqueous samples obtained with tetracycline enzyme-linked immunosorbent assay (ELISA) kit. Matrix Minnehaha Creek

Silver Lake

Soil saturation extracts Hubbard loamy sand Webster clay loam Seelyeville muck Rozetta silt loam

Spiked concentration of tetracycline

Dilution factor†

Mean tetracycline concentration using ELISA

Mean percent recovery

␮g L⫺1 0.15 0.45 1.35 4.05 0.15 0.45 1.35 4.05

2 2 2 2 2 2 2 2

␮g L⫺1 0.17 0.41 1.23 4.24 0.15 0.46 1.40 3.97

% 114 91 91 105 100 102 104 98

4.5 13.5 4.5 13.5 4.5 13.5 4.5 13.5

2 5 2 5 2 5 2 5

4.6 13.7 4.5 13.9 5.5 12.8 4.9 12.5

102 101 100 103 122 95 109 93

† The samples were diluted using dilution buffer supplied with the kit to bring the concentration within linear part of standard curve (0.15–4.05 ␮g L⫺1).

recovery of 101% (Table 4). There were three samples (out of 16) where tetracycline recovery was 91, 91, and 93%. These results show that these ELISA kits can be used to quantify tylosin and tetracycline antibiotics from aqueous samples with a variety of background matrices found in agricultural settings.

Cross-Reactivity of Tetracycline-Class Antibiotics The specificity of the tetracycline kit was also determined by analyzing the cross-reactivity of other tetracy-

clines such as chlortetracycline, oxytetracycline, demeclocycline, and doxycycline. The tetracycline ELISA test showed nearly 100% specificity for tetracycline and chlortetracycline (Table 5). However, the specificity of ELISA test kit for demecocycline was between 21 and 37%, for doxycycline between 15 and 27%, and for oxytetracycline between 7 and 20% (Table 5). This clearly shows that quantification of these tetracyclineclass antibiotics (oxytetracycline, doxycycline, and demecocycline) in aqueous samples is not possible using this ELISA kit.

Table 5. Detection of different tetracycline antibiotics using tetracycline enzyme-linked immunosorbent assay (ELISA). Matrix

Spiked concentrations of various antibiotics

Dilution factor†

Mean concentration using ELISA

Mean percent recovery

␮g L⫺1

%

4.9 10.1 20.5 5.2 10.2 19.6 1.1 3.7 6.4 0.8 1.5 4.6 0.6 0.7 1.6

98 101 102 104 102 98 22 37 32 15 15 23 12 7 8

5.2 9.7 20.1 4.7 9.9 20.7 1.3 2.1 4.3 0.9 2.1 5.3 1.0 1.2 2.0

104 97 101 94 99 104 26 21 22 18 21 27 20 12 10

␮g L⫺1 Standard buffer Tetracycline Chlortetracycline Demecocycline Doxycycline Oxytetracycline

5 10 20 5 10 20 5 10 20 5 10 20 5 10 20

2 5 10 2 5 10 2 5 10 2 5 10 2 5 10

5 10 20 5 10 20 5 10 20 5 10 20 5 10 20

2 5 10 2 5 10 2 5 10 2 5 10 2 5 10

Soil saturations extract Tetracycline Chlortetracycline Demecocycline Doxycycline Oxytetracycline

† The samples were diluted using dilution buffer supplied with the kit to bring the concentration within linear part of standard curve (0.15–4.05 ␮g L⫺1).

255

KUMAR ET AL.: ELISA ASSAY FOR ANTIBIOTIC DETERMINATION

Comparison of Enzyme-Linked Immunosorbent Assay and Liquid Chromatography– Mass Spectrometry Four liquid swine manure samples obtained from hog farmers using chlortetracycline and tylosin in swine feed were analyzed using both ELISA and LC–MS. The mass spectra of the peaks from the manure samples matched with the mass spectra of the standards (Fig. 2). Two out of four manure samples had both chlortetracycline and tylosin present; one manure sample contained only chlortetracycline and one sample contained only traces of tylosin (Table 6). The chlortetracycline and tylosin concentrations in manure samples using ELISA varied from 0 to 7931 and 0 to 5230 ␮g L⫺1, respectively. The corresponding values using LC–MS varied from 0 to 5230 and 0 to 3780 ␮g L⫺1, respectively.

DISCUSSION The present study shows that the antibiotic kits commercially available to quantify tylosin and tetracycline residues in meat, milk, and honey could be adapted for analysis of environmental waters. Although background matrices affected the measured optical densities of blank samples in relation to negative controls, the variation was reduced substantially by mixing samples with dilution buffers supplied with the kits (negative blank) in a 1:1 ratio. In addition, the variation due to time between tests and to different sample matrices was further reduced when OD values were normalized and expressed as inhibition values. These types of adjustments have been suggested by the makers of ELISA kits and have been commonly used by enzymologists (Poirier and Connor, 1982; Takatori, 1982). The ELISA tests were highly selective for the respective antibiotic. The recoveries of both tylosin and tetracycline in spiked samples at various concentrations and background matrices were close to 100%. These results show that these kits can be adapted to quantify tylosin and tetracycline concentrations in water and manure samples. Furthermore, the tetracycline ELISA test is specific for tetracycline and chlortetracycline only. Samples containing oxytetracycline, doxycycline, and demecocycline cannot be quantified using the tetracycline ELISA kit. In general, the ELISA values for chlortetracycline in manure samples were higher than values obtained using LC–MS analysis (Table 6). This may be due to the presence of transformed or decay products of chlortetracycline in manure that reacted with ELISA antibody thus resulting in higher concentrations of chlortetracycline as compared with LC–MS. The concentrations of tylosin in manure samples obtained with ELISA and LC–MS were comparable.

Fig. 2. Liquid chromatography–mass spectrometry (LC–MS) chromatograms of a standard mixture of chlortetracycline (m/z 479, RT ⫽ 7.8) and tylosin (m/z 916, RT ⫽ 12.9) compared with different swine manure samples.

the presence of antibiotics and transformed by-products in various water quality samples. Moreover, these tests require a small sample volume (⬍100 ␮L), are field portable, and work at very low but environmentally significant concentrations (⬎0.10 ␮g L⫺1). These abovereported ELISA procedures can provide a low-cost screening tool for the presence of tylosin, tetracycline, and chlortetracycline in surface and ground waters, runoff samples, and manure. Comparison of ELISA with LC–MS technique showed that although ELISA is a Table 6. Comparison of chlortetracycline and tylosin concentrations in swine manure samples using enzyme-linked immunosorbent assay (ELISA) and liquid chromatography–mass spectrometry (LC–MS). Chlortetracycline Sample

ELISA

CONCLUSIONS The results show that ELISA tests are inexpensive (approximately $5 per sample for tylosin and $15 per sample for tetracycline) and quick monitoring tools for

Tylosin

LC–MS

ELISA ␮g

Manure Manure Manure Manure

1 2 3 4

7931 0 5146 4698

5230 0 4310 3540

LC–MS

L⫺1 4032 6 0 3304

3780 0 0 3690

256

J. ENVIRON. QUAL., VOL. 33, JANUARY–FEBRUARY 2004

reliable technique to screen a large number of samples, some samples should be run through LC–MS to confirm the presence of a given antibiotic or its transformation product. ACKNOWLEDGMENTS This research was supported with funds from the U.S. Geological Survey through the National Institute of Water Resources; USDA Cooperative State Research, Education, and Extension Service through the National Center for Manure and Animal Waste Management; Minnesota Pork Producers Association; and the graduate school of the University of Minnesota. The authors are grateful to Professor Sagar Goyal for his comments on the manuscript. Constructive criticisms from Associate Editor Dr. Jorge S. Domingo and two anonymous reviewers have helped us improve this manuscript.

REFERENCES Aga, D.S., and E.M. Thurman. 1993. Coupling solid-phase extraction and enzyme-linked immunosorbent assay for ultratrace determination of herbicides in pristine water. Anal. Chem. 65:2894–2898. American Chemical Society Committee on Environmental Improvement and Subcommittee on Environmental Analytical Chemistry. 1980. Guidelines for data acquisition and data quality evaluation in environmental chemistry. Anal. Chem. 52:2242–2249. Church, D.C., and W.G. Pond. 1982. Basic animal nutrition and feeding. 2nd ed. John Wiley & Sons, New York. Cooper, A.D., G.W.F. Stubbings, M. Kelly, J.A. Tarbin, W.H.H. Farrington, and G. Shearer. 1998. Improved method for the on-line metal chelate affinity chromatography-high-performance liquid chromatographic determination of tetracycline antibiotics in animal products. J. Chromatogr. A 812:321–326. De Wasch, K., L. Okerman, S. Croubels, H. De Brabander, J. Van Hoof, and P. De Backer. 2001. Detection of residues of tetracycline antibiotics in pork and chicken meat: Correlation between results of screening and confirmatory tests. Analyst 123:2737–2741. Draisci, R., F.D. Quadri, L. Achene, G. Volpe, L. Palleschiand, and G. Palleschi. 2001. A new electrochemical enzyme-linked immunosorbent assay for the screening of macrolide antibiotic residues in bovine meat. Analyst 126:1942–1946. Environmental Media Services. 2000. Agricultural use of antibiotics [Online]. Available at http://www.ems.org/antibiotics/antibiotics_ food.html (verified 26 Sept. 2003). EMS, Washington, DC. Ford, T. 1994. Pollutant effects on the microbial ecosystem. Environ. Health Perspect. 102(Supplement 12):45–48. Goldburg, B. 1999. Antibiotic resistance: Grave threat to human health [Online]. Available at http://www.environmentaldefense. org/pubs/Newsletter/1999/Jun/j_antibi.htm (verified 26 Sept. 2003). Environ. Defense Fund, New York. Halling-Sørensen, B., S.N. Nielsen, P.F. Lanzky, F. Ingerslev, H.C.

Holten Lutzheft, and S.E. Jørgensen. 1998. Occurrence, fate and effects of pharmaceutical substances in the environment—A review. Chemosphere 36:357–393. Herron, P.R., I.K. Toth, G.H.J. Heiling, A.D.L. Akkermans, A. Karagouni, and E.M.H. Wellington. 1997. Selective effect of antibiotics on survival and gene transfer of streptomycetes in soil. Soil Biol. Biochem. 30:673–677. Jørgensen, S.E., and B. Halling-Sørensen. 2000. Drugs in the environment. Chemosphere 40:691–699. Koplin, D.W., E.T. Furlong, M.T. Meyer, E.M. Thurman, S.D. Zaugg, L.B. Barber, and H.T. Buxton. 2002. Pharmaceuticals, hormones, and other organic wastewater contaminants in US streams, 1999– 2002: A national reconnaissance. Environ. Sci. Technol. 36:1202–1211. Kumar, K., A. Thompson, A. Singh, and S.C. Gupta. 2002. Adsorption of antibiotics on soils [CD-ROM]. In 2002 Annual meetings abstracts. ASA, CSSA, and SSSA, Madison, WI. Lee, H.J., M.H. Lee, P.D. Ryu, H. Lee, and M.H. Cho. 2001. Enzymelinked immunosorbent assay for screening the plasma residues of tetracycline antibiotics in pigs. J. Vet. Med. Sci. 63:553–556. Levy, S.B. 1992. The antibiotic paradox: How miracle drugs are destroying the miracle. Plenum Publ., New York. Lindsey, M.E., M. Meyer, and E.M. Thurman. 2001. Analysis of trace levels of sulfonamide and tetracycline antimicrobials in groundwater and surface water using solid-phase extraction and liquid chromatography/mass spectrometry. Anal. Chem. 73:4640–4646. Loke, M.L., F. Ingerslev, B. Halling-Sorensen, and J. Tjornelund. 2000. Stability of tylosin A in manure containing test systems determined by high performance liquid chromatography. Chemosphere 40:759–765. Mascher, A., S. Lavagnoli, and M. Curatolo. 2001. Determination of residual oxytetracycline in honey with an immunoassay kit. Apidologie 27:229–233. Meyer, M.T., J.E. Bumgarner, J.L. Varns, J.V. Daughtridge, E.M. Thurman, and K.A. Hostetler. 2000. Use of radioimmunoassay as a screen for antibiotics in confined feeding operations and confirmation by liquid chromatography/mass spectrometry. Sci. Total Environ. 248:181–187. Poirier, M.L., and R.J. Connor. 1982. Radioimmunoassay for 2-acetylaminofluorene-DNA adducts. Methods Enzymol. 84:607–618. Sacher, F., F.T. Lange, H.-J. Brauch, and I. Blankenhorn. 2001. Pharmaceuticals in groundwaters: Analytical methods and results of a monitoring program in Baden-Wu¨rttemberg, Germany. J. Chromatogr. A 938:199–210. Takatori, T. 1982. Radioimmunoassay of pyrazolone derivatives. Methods Enzymol. 84:596–603. Thurman, E.M., M. Meyer, M. Pomes, C.A. Perry, and A.P. Schwab. 1990. Enzyme-linked immunosorbent assay compared with gas chromatography/mass spectrometry for the determination of triazine herbicides in water. Anal. Chem. 62:2043–2048. USEPA. 1992. EPA immunoassay informational meeting. Environ. Monitoring Systems Lab., Las Vegas, NV. Zhu, J., D.D. Snow, D.A. Cassada, S.J. Monson, and R.F. Spalding. 2001. Analysis of oxytetracycline, tetracycline, and chlortetracycline in water using solid-phase extraction and liquid chromatography-tandem mass spectrometry. J. Chromatogr. A 928:177–186.