Clinical Chemistry 47, No. 12, 2001

(63 ␮g/L). We added 300 ␮L of the patient’s serum to 600 ␮L of a suspension of protein A-Sepharose from Pharmacia (binding capacity, 16 mg human IgG/mL) and incubated the mixture with rotation at room temperature for 20 h. This treatment had no effect on the interference exhibited in the VIDAS assay. This finding suggested that the interference might be attributable to IgM antibodies. Finally, the serum was treated in heterophilic blocking tubes (HBTs) from Scantibodies, according to the manufacturer’s instructions. According to Scantibodies, these tubes contain a predispensed lyophilized heterophilic antibody-blocking active agent targeted to neutralize heterophilic antibody interference in immunoassays. This agent is a murine IgG directed against human IgM (9 ). After the treatment in HBTs, the prolactin concentration measured by the VIDAS decreased from ⬎200 ␮g/L to 17.4 ␮g/L, whereas the Elecsys result remained practically unchanged (23.3 vs 21.6 ␮g/L). The finding that the prolactin concentration measured by the VIDAS decreased after treatment with the HBT blocking agent supports the hypothesis that the interfering antibody was an IgM, an idiotypic antibody that binds to a unique idiotype on the VIDAS antibodies. This epitope was apparently not present on the antibodies used with the other assays or on the immunoglobulins probably incorporated in the VIDAS assay reagents as blocking agents. Another hypothesis might be that the Elecsys, Vitros, and AxSYM reagents contain the HBT blocking agent. The interfering antibody is likely to be a natural IgM idiotypic antibody (10 ). Unfortunately, no more serum was available to confirm the IgM nature of the interfering antibody. Macroprolactinemia is a frequent source of falsely increased prolactin values. Interference from antibodies is much less common, but may still occur with current prolactin IMAs. Among the four assays used in this study, three were efficiently protected against the antibodies contained in the patient’s serum. As described previously with human

chorionic gonadotropin assays (11 ), treatment with HBTs was able to obviate the interference. In similar cases of hyperprolactinemia without image of pituitary tumor, after macroprolactin screening prolactin should be assayed by two or more IMAs. The use of HBT blocking agents should also be encouraged to prevent unnecessary anxiety and costly medical procedures and to reinforce clinicians’ and patients’ confidence in diagnostic assays.

positive hCG results [Letter]. Clin Chem 2001; 47:1332–3.

Re´ my Sapin1* Chantal Simon2 1

Laboratoire Universitaire de Biophysique Unite´ d’Analyses Endocriniennes CNRS UPRES-A 7004 Universite´ Louis Pasteur Faculte´ de Me´decine 67085 Strasbourg Cedex, France 2

This study was supported in part by a grant from the Hoˆ pitaux Universitaires de Strasbourg. S. Doffoel and C. Bensimon are acknowledged for the Vitros ECi and AxSYM prolactin measurements, respectively. We thank N. Heider for carefully reviewing the English in the manuscript, bioMe´ rieux for providing the VIDAS analyzer and prolactin assay reagents, and Immunotech for providing the protein A suspension. References 1. St-Jean E, Blain F, Comtois R. High prolactin levels may be missed by immunoradiometric assay in patients with macroprolactinomas. Clin Endocrinol 1996;44:305–9. 2. Gibson G, Schmit P, Thix J, Hoffman JP, Humbel RL. Prolactin results for samples containing macroprolactin are method and sample dependent [Technical Brief]. Clin Chem 2001;47: 331–3. 3. Vandalem JL, Brakier T, Pirens G, Hennen G. Falsely increased values of pituitary hormones in a healthy woman due to interfering heterophilic immunoglobulins. Ann Endocrinol 1981; 42:539 – 40. 4. Dericks-Tan JS, Jost A, Schwedes U, Taubert HD. Pseudohypergonadotropinemia and pseudohyperprolactinemia induced by heterophilic antibodies? Klin Wochenschr 1984;62:265–73. 5. Jeffcoate WJ, Selby C, Powell RJ. RIA artifact associated with circulating anti-rabbit antibodies. Clin Endocrinol 1988;28:39 – 43. 6. Hellthaler G, Briedigkeit L, Zinn W. Erroneous prolactin determination caused by heterophile antibodies. Geburtshilfe Frauenheilkd 1995; 55:M55– 6. 7. Fahie-Wilson M, Brunsden P, Surrey J, Everitt A. Macroprolactin and the Roche Elecsys prolactin assay; characteristics of the reaction and detection by precipitation with polyethylene glycol [Technical Brief]. Clin Chem 2000;46:1993–5. 8. Levinson SS. Test interferences from endogenous antibodies. J Clin Ligand Assay 1997;20: 180 –9. 9. Kricka LJ. Human anti-animal antibody interferences in immunological assays [Review]. Clin Chem 1999;45:942–56. 10. Hurez V, Kaveri SV, Kazatchkine MD. Expression and control of the natural autoreactive IgG repertoire in normal serum. Eur J Immunol 1993;23:783–9. 11. Butler SA, Cole LA. Use of heterophilic antibody-blocking agent (HBT) in reducing false-

Service de Me´decine Interne et Nutrition Hoˆ pital de Hautepierre 67098 Strasbourg Cedex, France *Address correspondence to this author at: Institut de Physique Biologique, Faculte´ de Me´ decine, F-67085 Strasbourg Cedex, France. Fax 33-3-90-24-40-57; email [email protected].

Target Values and Method Evaluation in Proficiency Testing Programs

To the Editor: A recent opinion article (1 ) in this journal presented sound metrologic principles and a plan of action to improve analytical accuracy in medical laboratories. In addition to the use of Certified Reference Materials outlined by Mu¨ ller (1 ), use of proficiency testing and external quality-control programs can have a broader impact on interlaboratory comparability. Ideally, target values should be derived from reference methods and reference materials, but both are limited in their availability, and the mean (or other indicator of central tendency) of participant data is frequently used. This has been shown to provide a basis of comparison often comparable to reference methods (2 ). The means of subsets of methods (peer-group means) are often reported as target values, a practice usually ascribed to method-dependent behavior of proficiency test specimens and so-called matrix effects (3, 4 ); this use of multiple “true values” in proficiency testing has been criticized (5, 6 ).

2186

In the US, the CLIA’88 serves as the basis for any accredited external quality-control program. The current regulations (7 ) state that only when “a specific method’s results demonstrate bias that is not observed with actual patient specimens, as determined by a defensible scientific protocol, a comparative method or a method group (‘peer’ group) may be used”. Nonetheless, peer group evaluation is commonplace, perhaps reflecting potential for sanctions associated with failure in proficiency testing. Here we share two examples where this practice would have masked true analytical bias. Case 1. In a recent proficiency test, the results for thyroxine from a participant laboratory demonstrated a high bias relative to both the robust mean (8 ) of all participants and that of the instrument peer group (Fig. 1). The degree of the high bias was less noticeable within the peer group; had the results been evaluated against the peer group, a CLIA score of 100% would have been obtained. The results were not peer graded, and three of the five results fell out of the allowable range and produced an unsatisfactory score of 40% (Fig. 1A, specimens 1–3). In spite of the bias demonstrated by this method, other laboratories using it attained a CLIA score of 100% when evaluated against the overall range or against the peer group-specific range, had the latter been used. Examination of this failure indicated that the bias was attributable to the precipitation, over time, of magnetic particles in the reagent solution onto the side of the reagent container, producing a shift in reported values. Precipitation would not have occurred had the reagent been consumed within 24 h. The failing laboratory had a low test volume and stored the unused reagent at 4 °C between analyses. The identified source of the bias was verified by reproducing the high-bias phenomenon with the stored reagent and by eliminating the high bias using the stored reagent but mixing it just before use (Fig. 1). After this investigation, the manufacturer issued a service notice in February 2001 and initiated activities to improve the reagent quality.

Letters

mended reference interval; proficiency test results for the analyzer now appear consistent with all-participant mean values.

Fig. 1. Relative bias of thyroxine results for five proficiency test specimens. Relative bias (where 0.0 is a result equal to the target concentration and ⫹1.0 is a result equal to the upper acceptable CLIA limit) is shown, using the all-method mean (‚) and peer-group mean (f) without (A) and with (B) reagent mixing. Target concentrations for specimens 1–5 were 113, 142, 221, 76, and 31 nmol/L (8.8, 11.0, 17.2, 5.9, and 2.4 ␮g/dL), respectively. Peer-group target concentrations were 121, 160, 263, 80, and 36 nmol/L (9.4, 12.4, 20.4, 6.2, and 2.8 ␮g/dL), respectively.

Case 2. Calcium is a non-peer-graded analyte in our proficiency test program with an allowable limit of ⫾0.25 mmol/L (1.0 mg/dL) from the all-method mean value (7 ). Methods agree well, and the failure rate in the New York State program is low [in the year 2000, 43 results of 6185 (0.7%) were unsatisfactory]. Over the course of several test events, a negative bias was consistently present for a particular instrument model with the o-cresolphthalein complexone method. The reported calcium values were either at or below the lower acceptable limit. The results within this instrument group were similar, and had peer group-specific target values been used, laboratories using this analyzer would have been awarded passing scores. The manufacturer indicated that they accommodated the known bias by a commensurate downward adjustment of the reference interval [1.85– 2.30 mmol/L (7.4 –9.2 mg/dL)]. This range was lower than that typically used by clinical laboratories [2.10 –2.55 mmol/L (8.4 –10.2 mg/dL)] or found by reference methods (9 ). Further investigation by the manufacturer led to changes in the calcium calibrator set point values. The set point adjustment (⬃9%) produced an increase in the recom-

These two examples demonstrate positive outcomes that may occur by avoiding peer-group evaluation in external quality assessment programs. Problems were identified that affected the quality of patient testing, resolutions were found, and comparability among methods was advanced. This supports the view of Bu¨ ttner (5 ) that in “proficiency testing, so called ‘peer group mean values’ . . . do not lead to any improvement of the trueness or therefore of the comparability”. References 1. Mu¨ ller MM. Implementation of reference systems in laboratory medicine. Clin Chem 2000; 46:1907–9. 2. Hartmann AE, Naito HK, Burnett RW, Welch MJ. Accuracy of participant results used as target values in the CAP Chemistry Survey Program. Arch Pathol Lab Med 1985;109:894 –903. 3. Rej, R. Proficiency testing, matrix effects, and method evaluation. Clin Chem 1994;40:345– 6. 4. Rautela GS. A manufacturer’s approach to development of matrix robust methods. Arch Pathol Lab Med 1993;117:691–5. 5. Bu¨ ttner J. The need for accuracy in laboratory medicine. Eur J Clin Chem Clin Biochem 1995; 33:981– 8. 6. Tietz NW, Rodgerson DO, Laessig RH. Are clinical laboratory proficiency tests as good as they can be? Clin Chem 1992;38:473–5. 7. US Department of Health and Human Services. Medicare, Medicaid and CLIA programs: regulations implementing the Clinical Laboratory Improvement Amendments of 1988 (CLIA), final rule. Fed Regist 1992;57:7002–186. 8. Paulson AS, Nicklin EH. Integrated distance estimators for linear models applied to some published data sets. Appl Stat 1983;32:32–50. 9. Pybus J, Feldman FJ, Bowers GN Jr. Measurement of total calcium in serum by atomic absorption spectrophotometry, with use of a strontium internal reference. Clin Chem 1970;16:998 –1007.

Robert Rej1,2* Carol S. Norton-Wenzel1 (Tim) Zhimin Cao1 1

Wadsworth Center for Laboratories and Research New York State Department of Health Albany, NY 12201 2

Department of Biomedical Sciences School of Public Health State University of New York Albany, NY 12201