VIEWPOINT

Where are we, and where are we heading in the device management of ventricular tachycardia/ventricular fibrillation? Rahul Mehra, PhD, Paul DeGroot, MS From Medtronic, Inc., Minneapolis, Minnesota. Cardiovascular disease is a leading cause of global mortality, accounting for almost 30% (17.5 million) of all deaths in the year 2005.1 It is estimated that approximately 20% to 40% of all cardiovascular deaths are sudden cardiac deaths (SCDs), and the majority are caused by ventricular tachyarrhythmias. The following sections discuss “where we are” and “where we are heading” in device management of ventricular tachyarrhythmias (ventricular tachycardia [VT]/ventricular fibrillation [VF]).

Where are we? Device management of spontaneous tachyarrhythmias dates to the early 1970s, when pacemakers were adapted to treat supraventricular tachycardias (SVTs).2 Compelled by the loss of a friend to SCD, Dr. Mirowski worked with engineers who eventually developed the first commercial implantable cardioverter-defibrillator (ICD) in the mid-1980s.3 First-generation ICDs were simple “shock-only devices” without pacing for bradycardia or antitachycardia pacing (ATP). Shocks were delivered via epicardial patches sewn onto the heart during thoracotomy. It was these simple devices that established the genesis of today’s ICD industry. Several significant technological improvements coupled with the clinical outcome data from trials allowed the ICD to become a frontline treatment for SCD survivors and patients with recurring VT/VF. First, the incorporation of a biphasic shock waveform significantly improved shock efficacy. This, in turn, paved the way for the use of transvenous leads. Finally, advances in ICD design allowed a significant reduction in device size, from approximately 110 to 35 cc. Also, having established a reliable safety net, ICDs now combine VF shock therapy with ATP to treat monomorphic VT, the predominant ventricular tachyarrhythmia. Finally, arrhythmia detection has advanced from being solely based on ventricular rate to incorporating sophisticated dual-chamber algorithms including electrogram morphology information to discriminate SVT from VT/VF. With these advances, annual ICD implantation rates in the United States increased from 9,000 in 1990 to 143,000 in 2005. Address reprint requests and correspondence: Dr. Rahul Mehra, Medtronic, Inc., 7000 Central Avenue NE, Mailstop T203, Minneapolis, Minnesota 55432-3576. E-mail address: [email protected].

Significant capabilities of modern ICDs include monitoring and diagnostic features for cardiac disease management. Tachyarrhythmia episode counters, heart rate patterns, and stored electrograms as well as self-diagnostic information such as battery status, capacitor charge time, and lead integrity greatly enhanced the clinicians’ ability to accurately determine the rhythm triggering device therapy as well as to identify potential problems with the ICD system.4 In addition to providing therapy for sustained ventricular tachyarrhythmias, newer devices are capable of providing information to assist in the management of comorbidities common in ICD recipients. Heart rate variability and atrial fibrillation rhythm and rate control parameters have provided some value in managing patients and predicting clinical outcomes.5–7 For example, in patients receiving cardiac resynchronization pacing therapy, heart rate variability has been shown to be lower in those with high risk for mortality and hospitalization, including a decrease in heart rate variability prior to hospitalization. With increasing use of sensors and memory capabilities, the number of diagnostic parameters will continue to grow. Strategies for communicating critically important information have also evolved with time. In the earliest generation of devices, information could be downloaded only during the follow-up visit. In the next generation, an auditory alarm could alert patients to device malfunctions. The patient, in turn, would communicate with the medical caregiver regarding these alerts. In the most recent devices, critically important information is transmitted automatically from a home-based unit to the physician or to a call center. Full interrogation of devices can be conducted by accessing a secure web site. The information is transmitted using short-distance telemetry from the implanted device to the base unit located at home and then via the Internet. Some devices can communicate information via a cellular phone, and this mobile technology allows communication of urgent information even when the patient is not at home. It will become increasingly important to demonstrate the value of these diagnostics to ensure that patient care is being optimized and that the caregiver is not burdened with irrelevant information. In addition to providing therapy for ventricular tachyarrhythmias, ICDs now provide a host of features that specifically target management of a patient’s heart failure status.

1547-5271/$ -see front matter © 2007 Heart Rhythm Society. All rights reserved.

doi:10.1016/j.hrthm.2006.09.010

100 Data from the DAVID (Dual Chamber and VVI Implantable Defibrillator) trial highlighted an unanticipated adverse cardiovascular outcome with unnecessary long-term right ventricular apical pacing.8 Thus, today’s devices include algorithms to minimize ventricular pacing in patients who historically would have been paced with conventional dualchamber pacing. A biventricular pacing feature is now used in ICDs for many heart failure patients in the wake of trials that demonstrated improvement in New York Heart Association class, quality of life, and mortality in patients receiving cardiac resynchronization therapy. ICDs also have the capability of measuring thoracic impedance to monitor pulmonary edema. Preliminary data have shown its potential for predicting impending decompensation from fluid overload.9 Finally, the COMPASS-HF (Chronicle Offers Management to Patients with Advanced Signs and Symptoms of Heart Failure) trial showed that continuous monitoring of intracardiac pressure improved outcomes in patients with chronic heart failure.10 This feature currently is incorporated into an ICD undergoing clinical investigation. As with other device features, the ability to continuously track fluid volume via pressure sensors could offer an advantage over intermittent recording of patient status by acute right heart catheterization11 or by variables such as body weight or B-type natriuretic peptide.12 However, prospective comparison of continuous intracardiac pressures and optimized home care, including frequent transmissions of weights, has not been performed. Diagnostic information as well as algorithms for termination and prevention of atrial tachyarrhythmias have been included in certain ICDs.7,13 These features may provide outcome benefit in a subset of patients (e.g., patients with atrial fibrillation and congestive heart failure) but require evaluation in prospective trials.

Indications The indications for ICD placement have undergone a major evolution. Early ICDs were narrowly indicated for SCD survivors or patients resistant to antiarrhythmic drug therapy. In 1997, the AVID (Antiarrhythmics Versus Implantable Defibrillators) trial established the superiority of ICDs over antiarrhythmic drugs for secondary prevention of SCD in patients with coronary artery disease and low ejection fraction.14 Not long after, multiple studies, including MADIT (Multicenter Automatic Defibrillator Implantation Trial), MADIT-II, and MUSTT (Multicenter Unsustained Tachycardia Trial), established the value of ICDs for the primary prevention of SCD.15–17 Most recently, SCD-HeFT (Sudden Cardiac Death–Heart Failure Trial) demonstrated the superiority of ICDs over amiodarone for ischemic and nonischemic patients with congestive heart failure and low ejection fraction.18 As a result of these series of trials, the number of patients in the United States who potentially could benefit from ICD has grown nearly 4.5-fold, from 270,000 to 1.3 million. In spite of these advances, patients with ICDs still die, a proportion of which are sudden deaths.19 Clearly, the ability of ICDs to prevent SCD is not

Heart Rhythm, Vol 4, No 1, January 2007 infallible. In some patients, defibrillation may not be possible, or the thresholds may be extremely high under extreme settings of metabolic imbalance, ischemia, and/or acute myocardial infarction.20 Electromechanical dissociation is a common cause of sudden death in ICD recipients and accounted for 45% of such deaths in one study.19

Challenges to increase acceptance of ICDs Today, fewer than 20% of patients in countries with developed economies who have indications for an ICD receive the device. There are several key challenges for increasing the acceptance of ICD therapy. The first challenge is to further determine the subgroup of patients who can most benefit from this technology and therefore improve the overall cost-effectiveness. Cost-effectiveness varies significantly depending on the patient receiving the device. Most trials to date have focused on the highest-risk subgroups because they yield event rates sufficiently high to design trials with reasonably small numbers of patients. These subgroups constitute only approximately 30% of all patients at risk for SCD.21 The lifetime cost-effectiveness of ICD therapy in this subgroup can vary from $40,000 to $80,000 per life-year saved, a value that is within the range of other medical therapies accepted in developed economies.22,23 However, improvements in device longevity, cost of implant, risk stratification, and long-term overall cost of device therapy can further improve cost-effectiveness. Better understanding of the patient’s needs allows physicians to use devices of varying prices today. The latest devices with the most desirable features tend to have higher prices because of the cost of research and development in a rapidly evolving technology. The second challenge is to improve the technology, facilitating implant and follow-up from the physician’s perspective and lowering morbidity from the patient’s perspective. Patients want ICDs that are smaller, result in fewer shocks, have less pain associated with defibrillation, require minimal follow-up, and last a long time so that the devices do not require replacement. The third challenge is to increase the awareness of this therapy among physicians and patients. The fourth issue and the one that has received much attention lately is ensuring product reliability and open communication with physicians and patients. If patients are to maintain trust in these lifesaving products, it is critically important that all companies have processes incorporated into the design and development to maintain the highest quality. All major manufacturers have reported pacemaker/defibrillator advisories associated with a risk of device malfunction. The manner in which this information is communicated to the customers is critically important. The Heart Rhythm Society recently made recommendations on communication to physicians and patients regarding device malfunction and corrective actions.24 Many of these recommendations have been standard practice at some companies. We need to address all of these challenges if im-

Mehra and DeGroot

Device Management of Ventricular Tachyarrhythmias

plantable defibrillators are to make a significant impact in reducing the incidence of SCD worldwide.

Where are we heading? One of the key technological challenges today is to reduce the number of inappropriate shocks delivered by ICDs, most of which result because of supraventricular tachyarrhythmias. Detection algorithms discriminate ventricular from supraventricular tachyarrhythmias by using timing as well as morphology of electrical cardiac signals. Although these algorithms will continue to improve, an alternative strategy would be to deliver therapy based on hemodynamic status. For example, measurement of intraventricular blood pressure using sensors could determine the urgency of terminating the tachyarrhythmia, irrespective of whether it is ventricular or supraventricular in origin.25 Acute animal experiments with a pressure sensor on a right ventricular defibrillation lead have shown promising preliminary results but require further testing.26 To date, right ventricular pressure sensor leads have undergone successful chronic clinical testing for heart failure management but not on a defibrillation lead. Another key issue has been the discomfort associated with shocks. Even if defibrillation thresholds are reduced 15% to 20% by optimizing electrodes and waveforms, they are unlikely to impact the shock discomfort. Fortunately, improvements in ATP have significantly reduced the need for shocks. There is also a need to make the devices easier to program and to reduce the time and cost of follow-up. The amount of data stored in the devices is increasing rapidly, resulting in an even greater need to communicate relevant information in a timely and efficient manner to the caregiver and the patient. Remote programming of devices should be possible in the future. The two driving forces of patient empowerment and the need to minimize the time required

101

by physicians to follow their patients could have a significant impact on the technology. The size of these devices will continue to decrease as the energy density of batteries and capacitors increases. In the future, it may be possible to prevent a certain proportion of ventricular tachyarrhythmias. Although the event that precipitates a tachyarrhythmia is not well understood, acute ischemia, autonomic or physical stress, or electrolyte disturbances likely play an important role. In ambulatory recording of spontaneous VT/VF, acute ST elevation preceding VT/VF is sometimes observed.27 Could precursors of these events be detected with implanted sensors and an antithrombotic or anti-ischemic agent delivered through an integrated drug pump for such episodes? Will this be an effective strategy for preventing VT/VF? The technology for delivering agents through implantable pumps exists today. However, they must be miniaturized further to provide devices that can prevent ventricular tachyarrhythmias. Sensitive and specific sensors for detection of ST segment also are needed. These are research challenges that require investigation. Even if we cannot prevent the first VT/VF event, preventing a cluster of VT/VF events by pacing algorithms may be possible. Lead technology will continue to improve with smaller leads that are less likely to develop fibrosis. Defibrillators without intracardiac leads that use subcutaneous electrodes for detection and defibrillation are presently being developed. These devices eliminate the morbidity associated with intracardiac leads. They require a subcutaneous electrode and have higher defibrillation energy requirements. In addition, they cannot deliver ATP or longterm antibradycardia pacing because the high energy levels required for pacing would be quite painful to the patient. Such devices are projected to undergo clinical testing soon.

Figure 1 Implantable defibrillators: key developments. ➢Shock-only defibrillator with epicardial leads ➢Transvenous leads ➢Biphasic shocks ➢“Device can” as active electrode ➢Size reduction ➢Dual-chamber detection ➢Atrial therapies ➢Improved ventricular antitachycardia pacing ➢Extensive diagnostics and monitoring ➢Minimizing ventricular pacing ➢Biventricular pacing for heart failure ➢Remote monitoring ➢Sensors for pulmonary edema ➢Intracardiac pressure monitoring for heart failure Future? ➢Hemodynamic sensors for reducing false-positive shocks ➢Artificial intelligence systems and autoprogramming ➢Prevention of ventricular tachycardia/ventricular fibrillation ➢Implantable cardioverter-defibrillator without intracardiac leads ➢Monitoring and management of comorbidities (e.g., ischemia, angina, hypertension, diabetes)

102 Over the last decade, ICDs have evolved from being primarily a defibrillator to being a cardiovascular disease management device for the treatment and monitoring of comorbidities such as atrial fibrillation and heart failure. This process of evolution is likely to continue, especially with monitoring and therapy for heart failure. Many of these patients also suffer from hypertension, diabetes, and angina. Monitoring and management of these comorbidities with appropriate sensors could become the next key development in the evolution of such devices (Figure 1).

Future Indications One challenge confronting us today is the risk stratification of patients to ensure that only those patients who would benefit most from a device receive one, thus improving cost-effectiveness. The patients most vulnerable to VT/VF are those with an arrhythmogenic substrate and triggers. Risk stratifiers such as ejection fraction, interventricular conduction delay, T-wave alternans, signal-averaged late potentials, heart rate variability, and heart rate turbulence reflect the status of the substrate. More recently, clinical studies of T-wave alternans have shown significant promise.28 The trigger can be an acute ischemic episode and/or runs of premature ventricular complexes, nonsustained VT, or sinus tachycardia possibly caused by autonomic disturbances. Genetic and protein markers possibly can be used to identify high-risk patients. The specificity of these risk stratifiers in predicting those patients most likely to develop VT/VF will be evaluated as large trials are conducted. One of the key limitations of many risk stratifiers is the static nature of the measurements. Measurements are made at one time during rest or exercise. In the future, it will be possible to make some of these measurements dynamically over longer periods using an implanted device. For example, having such algorithms in a pacemaker or implanting a small monitoring device could allow monitoring of these parameters in the ambulatory setting in order to risk stratify patients. In such a case, heart rate variability, T-wave alternans, etc., can be measured over long periods under various autonomic stressors. These strategies will allow more widespread use of this technology with a better ability to match this technology to those patients most likely to benefit. This should help reduce further the incidence of SCD. However, at present, evidence-based medicine dictates that patients who meet currently accepted indications according to established guidelines should receive this potentially lifesaving therapy.

References 1. World Health Statistics 2006. WHO Press, 2006. Geneva, Switzerland. 2. Ryan GF, Easley RM, Zaroff LI, Goldstein S. Paradoxical use of a demand pacemaker in treatment of supraventricular tachycardia due to the WolffParkinson-White syndrome. Circulation 1968;38:1037–1043. 3. Mirowski M, Reid PR, Mower MM, Watkins L, Gott VL, Schauble JF, Langer A, Heilman MS, Kolenik SA, Fischell RE, Weisfeldt ML. Termination of malignant ventricular arrhythmias with an implanted automatic defibrillator in human beings. N Engl J Med 1980;303:322–324. 4. Sarter BH, Callans DJ, Gottlieb GD, Schwartzman DS, Marchlinski FE. Implantable defibrillator diagnostic storage capabilities: evolution, current status, and future utilization. Pacing Clin Electrophysiol 1998;21:1287–1298.

Heart Rhythm, Vol 4, No 1, January 2007 5. Adamson PB, Smith AL, Abraham WT, Kleckner KJ, Stadler RW, Shih A, Rhodes MM. Continuous autonomic assessment in patients with symptomatic heart failure: prognostic value of heart rate variability measured by an implanted cardiac resynchronization device. Circulation 2004;110:2389 –2394. 6. Glotzer TV, Hellkamp AS, Zimmerman J, Sweeney MO, Yee R, Marinchak R, Cook J, Paraschos A, Love J, Radoslovich G, Lee KL, Lamas GA. Atrial high rate episodes detected by pacemaker diagnostics predict death and stroke: report of the Atrial Diagnostics Ancillary Study of the Mode Selection Trial (MOST). Circulation 2003;107:1614 –1619. 7. Mehra R, Ziegler P, Koehler J. Diagnostic information in implantable devices that pertain to endpoints in atrial fibrillation studies. Heart Rhythm 2004;1:B64 –B69. 8. Wilkoff BL, Cook JR, Epstein AE, Greene HL, Hallstrom AP, Hsia H, Kutalek SP, Sharma A, Dual Chamber and VVI Implantable Defibrillator Trial Investigators. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial. JAMA 2002;288:3115–3123. 9. Yu CM, Wang L, Chau E, Chan RH, Kong SL, Tang MO, Christensen J, Stadler RW, Lau CP. Intrathoracic impedance monitoring in patients with heart failure: correlation with fluid status and feasibility of early warning preceding hospitalization. Circulation 2005;112:841– 848. 10. Bourge RC, Aranda JM, Aaron MF, Abraham WT, Magalski A, Smith AL, Roettiger AD, Sparks B, Tallaj J. Clinical outcome of a heart failure management strategy based on continuous hemodynamic monitoring (abstr). Circulation 2005;112(Suppl II):II-639. 11. Adamson PB, Kjellström B, Braunschweig F, Magalski A, Linde C, Kolodiezj A, Cremers B, Bennett T. Ambulatory hemodynamic monitoring from an implanted device: components of continuous 24-hour pressures that correlate to supine resting conditions and acute right heart catheterization. Congest Heart Fail 2006;12:14 –19. 12. Lewin L, Ledwidge M, O’Loughlin C, McNally C, McDonald K. Clinical deterioration in established heart failure: What is the value of BNP and weight gain in aiding diagnosis? Eur J Heart Fail 2005;7:953–957. 13. Mehra R, Euler D, Musley S, Warman E. Management of atrial fibrillation patients with dual defibrillators. In: Israel CW, Barold SS, editors. Advance in the treatment of tachyarrhythmias: pacing, cardioversion, and defibrillation. Armonk, NY: Futura Publishing Co., 2002: 401– 418. 14. Antiarrhythmics Versus Implantable Defibrillators (AVID) Investigators. A comparison of antiarrhythmic therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. N Engl J Med 1997;33: 1576 –1583. 15. Moss A, Hall W, Cannom D, Daubert J, Higgins S, Klein H, Levine J, Saksena S, Waldo A, Wilber D, Brown M, Heo M, MADIT Investigators. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. N Engl J Med 1996;335:1933–1940. 16. Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, Daubert JP, Higgins SL, Brown MW, Andrews ML, MADIT II Investigators. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med 2002;346:877– 883. 17. Buxton AE, Lee KL, Fisher JD, Josephson ME, Prystowsky EN, Hafley G. Randomized study of the prevention of sudden death in patients with coronary artery disease. N Engl J Med 1999;341:1882–1890. 18. Bardy GH, Lee KL, Mark DB, Poole JE, Packer DL, Boineau R, Domanski M, Troutman C, Anderson J, Johnson G, McNulty SE, Clapp Channing N, Davidson-Ray LD, Fraulo ES, Fishbein DP, Luceri RM, Ip JH, SCD-HeFT Investigators. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 2005;352:225–237. 19. Mitchell LB, Pineda EA, Titus JL, Bartosch PM, Benditt DG. Sudden death in patients with implantable cardioverter defibrillators: the importance of postshock electromechanical dissociation. J Am Coll Cardiol 2002;39:1323–1328. 20. Anastasiou-Nana MI, Tsagalou EP, Charitos C, Siafakas KX, Drakos S, Terrovitis JV, Ntalianis A, Doufas A, Mavrikakis J, Nanas JN. Effects of transient myocardial ischemia on the ventricular defibrillation threshold. Pacing Clin Electrophysiol 2005;28:97–101. 21. Myerburg RJ, Mitrani R, Interian A, Castellanos A. Interpretation of outcomes of antiarrhythmic clinical trials. Design features and population impact. Circulation 1998;97:1514 –1521. 22. Zwazinger J, Hall WJ, Dick AW, Zhao H, Mushlin AI, Hahn RM, Wang H, Andrews ML, Mooney C, Wang H, Moss AJ. The cost-effectiveness of implantable cardioverter-defibrillators: results from the Multicenter Automatic Defibrillator Implantation Trial (MADIT)-II . J Am Coll Cardiol 2006;47:2310 –2318. 23. Mark DB, Nelson CL, Anstrom KJ, Al-Khatib SM, Tsiatis AA, Cowper PA, Clapp-Channing NE, Davidson-Ray L, Poole JE, Johnson G, Anderson J, Lee KL, Bardy GH, SCD-HeFT Investigators. Cost-effectiveness of defibrillator therapy or amiodarone in chronic stable heart failure: results from the Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT). Circulation 2006;114:135–142.

Mehra and DeGroot

Device Management of Ventricular Tachyarrhythmias

24. Draft recommendations report by the Heart Rhythm Society Task Force on Device performance. Policies and guidelines 2006. Available at: Heartrhythm.org. 25. Hegbom F, Hoff PI, Oie B, Folling M, Zeijlemaker V, Lindemans F, Ohm OJ. RV function in stable and unstable VT: is there a need for hemodynamic monitoring in future defibrillators? Pacing Clinical Electrophysiol 2001;24:172–182. 26. Kremers MS, Bailey JR, Wheelan KR, Franklin JO, Johnson WB, Mc Henry B, Sakowski B, Zmijewski M, Whitman T. Acute implant experience with a novel right ventricular pressure sensor defibrillation lead. Heart Rhythm 2005;2(1S):S5.

103

27. LeClercq JF, Coumel P. Lessons from recordings of sudden death by Holter monitoring. In: De Luna AB, Brugada P, Aguilar JC, Navarro-Lopez F, editors. Sudden cardiac death. Norwell, MA: Kluwer Academic Publishers, 1991: 99 –113. 28. Bloomfield DM, Bigger JT, Steinman RC, Namerow PB, Parides MK, Curtis AB, Kaufman ES, Davidenko JM, Shinn TS, Fontaine JM. Microvolt T-wave alternans and risk of death or sustained ventricular arrhythmias in patients with left ventricular dysfunction. J Am Coll Cardiol 2006;47:456 – 463.